Improved liver fat and R2* quantification at 0. 55 T using locally low‐rank denoising
Magnetic Resonance in Medicine · 2025
DOI: 10.1002/mrm.30324 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2025. Verfasst von Shu-Fu Shih und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiological Sciences, University of California Los Angeles, Los Angeles, California, USA. Hauptthemen laut MEDLINE: Phantoms, Imaging, Liver, Monte Carlo Method, Magnetic Resonance Imaging. Die Arbeit wurde bisher 10-mal zitiert (Crossref) und nennt 54 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: NIDDK NIH HHS, NIBIB NIH HHS, National Science Foundation.
Abstract (Originalsprache)
Purpose
To improve liver proton density fat fraction (PDFF) and quantification at 0.55 T by systematically validating the acquisition parameter choices and investigating the performance of locally low‐rank denoising methods.
Methods
A Monte Carlo simulation was conducted to design a protocol for PDFF and mapping at 0.55 T. Using this proposed protocol, we investigated the performance of robust locally low‐rank (RLLR) and random matrix theory (RMT) denoising. In a reference phantom, we assessed quantification accuracy (concordance correlation coefficient [] vs. reference values) and precision (using SD) across scan repetitions. We performed in vivo liver scans (11 subjects) and used regions of interest to compare means and SDs of PDFF and measurements. Kruskal–Wallis and Wilcoxon signed‐rank tests were performed ( p < 0.05 considered significant).
Results
In the phantom, RLLR and RMT denoising improved accuracy in PDFF and with >0.992 and improved precision with >67% decrease in SD across 50 scan repetitions versus conventional reconstruction (i.e., no denoising). For in vivo liver scans, the mean PDFF and mean were not significantly different between the three methods (conventional reconstruction; RLLR and RMT denoising). Without denoising, the SDs of PDFF and were 8.80% and 14.17 s −1 . RLLR denoising significantly reduced the values to 1.79% and 5.31 s −1 ( p < 0.001); RMT denoising significantly reduced the values to 2.00% and 4.81 s −1 ( p < 0.001).
Conclusion
We validated an acquisition protocol for improved PDFF and quantification at 0.55 T. Both RLLR and RMT denoising improved the accuracy and precision of PDFF and measurements.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiological Sciences, University of California Los Angeles, Los Angeles, California, USA.
- Department of Bioengineering, University of California Los Angeles, Los Angeles, California, USA.
- Ming Hsieh Department of Electrical and Computer Engineering, University of Southern California, Los Angeles, California, USA.
- MR R&D Collaborations, Siemens Medical Solutions USA, Inc., Los Angeles, California, USA.
ORCID-Profile: Shu-Fu Shih, Bilal Tasdelen, Ecrin Yagiz, Zhaohuan Zhang, Xiaodong Zhong, Sophia X Cui, Krishna S Nayak, Holden H Wu
Förderung
- NIDDK NIH HHS
- NIBIB NIH HHS
- National Science Foundation
- National Institute of Biomedical Imaging and Bioengineering
- National Institute of Diabetes and Digestive and Kidney Diseases
Zitieren
APA
Shih, S. F., Tasdelen, B., Yagiz, E., Zhang, Z., Zhong, X., Cui, S. X., Nayak, K. S., & Wu, H. H. (2025). Improved liver fat and R2* quantification at 0. 55 T using locally low‐rank denoising. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.30324
Vancouver
Shih SF, Tasdelen B, Yagiz E, Zhang Z, Zhong X, Cui SX, et al. Improved liver fat and R2* quantification at 0. 55 T using locally low‐rank denoising. Magnetic Resonance in Medicine. 2025. doi:10.1002/mrm.30324
BibTeX
@article{Shih2025,
author = {Shih, Shu-Fu and Tasdelen, Bilal and Yagiz, Ecrin and Zhang, Zhaohuan and Zhong, Xiaodong and Cui, Sophia X and Nayak, Krishna S and Wu, Holden H},
title = {Improved liver fat and R2* quantification at 0. 55 T using locally low‐rank denoising},
journal = {Magnetic Resonance in Medicine},
year = {2025},
doi = {10.1002/mrm.30324}
}
EUVIMED