Development of an algorithm for proton dose calculation in magnetic fields
Medical Physics · 2024
DOI: 10.1002/mp.17262 ↗Kurz gefasst
Fachartikel in der Zeitschrift Medical Physics, 2024. Verfasst von Yue Gu und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Engineering and Applied Physics, University of Science and Technology of China, Hefei, Anhui, China. Hauptthemen laut MEDLINE: Algorithms, Magnetic Fields, Proton Therapy, Radiotherapy Dosage. Die Arbeit nennt 39 Literaturangaben. Gefördert von: Ministry of Science and Technology of China Key Research and Development Projects, Fundamental Research Funds for the Central Universities, Hefei Comprehensive National Science Center.
Abstract (Originalsprache)
Background
The advantages of proton therapy can be further enhanced with online magnetic resonance imaging (MRI) guidance. One of the challenges in the realization of MRI‐guided proton therapy (MRPT) is accurately calculating the radiation dose in the presence of magnetic fields.
Purpose
This study aims to develop an efficient and accurate proton dose calculation algorithm adapted to the presence of magnetic fields.
Methods
An analytical‐numerical radiation dose calculation algorithm, Proton and Ion Dose Engine (PRIDE), was developed. The algorithm combines the pencil beam algorithm (PBA) with a novel iterative voxel‐based ray‐tracing algorithm. The new ray‐tracing method uses fewer assumptions and ensures broader applicability for proton beam trajectory prediction in magnetic fields, and has been compared to Wolf's method and Schellhammer's method. The accuracy of PRIDE algorithm was validated on three phantoms and two practical plans (one single‐field water plan and one prostate tumor plan) in different magnetic field strengths up to 3.0 T. The validation was performed by comparing the results against the Monte Carlo (MC) simulations, using the global gamma index criteria of 2%/2 mm and 3%/3 mm with a 10% threshold.
Results
PRIDE showed good agreement with MC in homogeneous and slab heterogeneous phantom, achieving gamma passing rates (%GPs) above 99% for 2%/2 mm criteria when magnetic field strength is not greater than 1.5 T. Although the agreement decreased for scenarios involving high proton energy (240 MeV) and strong magnetic field (3.0 T), the 2%/2 mm %GPs still remained above 98%. In lateral heterogeneous phantom, the accuracy of PRIDE decreased due to the PBA's limitation. For the two practical plans in different magnetic fields, %GPs exceeded 98% and 99% for 2%/2 mm and 3%/3 mm criteria, respectively.
Conclusions
PRIDE can perform efficient and accurate proton dose calculation in magnetic fields up to 3.0 T, and is expected to work as a useful tool for proton dose calculation in MRPT.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Engineering and Applied Physics, University of Science and Technology of China, Hefei, Anhui, China.
- Hefei Ion Medical Center, The First Affiliated Hospital of USTC, Division of Life Sciences and Medicine, University of Science and Technology of China, Hefei, Anhui, China.
- Ion Medical Research Institute, University of Science and Technology of China, Hefei, Anhui, China.
- Department of Radiation Oncology, the First Affiliated Hospital of USTC, Division of Life Sciences and Medicine, University of Science and Technology of China, Hefei, Anhui, China.
Förderung
- Ministry of Science and Technology of China Key Research and Development Projects
- Fundamental Research Funds for the Central Universities
- Hefei Comprehensive National Science Center
Zitieren
APA
Gu, Y., Wang, Y., Liu, M., Lu, H. M., & Yang, Y. (2024). Development of an algorithm for proton dose calculation in magnetic fields. Medical Physics. https://doi.org/10.1002/mp.17262
Vancouver
Gu Y, Wang Y, Liu M, Lu HM, Yang Y. Development of an algorithm for proton dose calculation in magnetic fields. Medical Physics. 2024. doi:10.1002/mp.17262
BibTeX
@article{Gu2024,
author = {Gu, Yue and Wang, Yuxiang and Liu, Meiqi and Lu, Hsiao-Ming and Yang, Yidong},
title = {Development of an algorithm for proton dose calculation in magnetic fields},
journal = {Medical Physics},
year = {2024},
doi = {10.1002/mp.17262}
}
EUVIMED