Time‐division multiplexing ( TDM ) sequence removes bias in T 2 estimation and relaxation‐diffusion measurements
Magnetic Resonance in Medicine · 13.08.2024
DOI: 10.1002/mrm.30246 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, online erschienen am 13.08.2024. Verfasst von Qiang Liu und 11 weiteren Autor:innen. Beteiligte Einrichtung: Brigham and Women's Hospital, Harvard Medical School, Boston, Massachusetts, USA. Hauptthemen laut MEDLINE: Phantoms, Imaging, Diffusion Magnetic Resonance Imaging, Brain, Algorithms. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 57 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: NIMH NIH HHS, NINDS NIH HHS, NIBIB NIH HHS.
Abstract (Originalsprache)
Purpose
To compare the performance of multi‐echo (ME) and time‐division multiplexing (TDM) sequences for accelerated relaxation‐diffusion MRI (rdMRI) acquisition and to examine their reliability in estimating accurate rdMRI microstructure measures.
Method
The ME, TDM, and the reference single‐echo (SE) sequences with six TEs were implemented using Pulseq with single‐band (SB) and multi‐band 2 (MB2) acceleration factors. On a diffusion phantom, the image intensities of the three sequences were compared, and the differences were quantified using the normalized RMS error (NRMSE). Shinnar–Le Roux (SLR) pulses were implemented for the SB‐ME and SB‐SE sequences to investigate the impact of slice profiles on ME sequences. For the in‐vivo brain scan, besides the image intensity comparison and T 2 ‐estimates, different methods were used to assess sequence‐related effects on microstructure estimation, including the relaxation diffusion imaging moment (REDIM) and the maximum‐entropy relaxation diffusion distribution (MaxEnt‐RDD).
Results
TDM performance was similar to the gold standard SE acquisition, whereas ME showed greater biases (3–4× larger NRMSEs for phantom, 2× for in‐vivo). T 2 values obtained from TDM closely matched SE, whereas ME sequences underestimated the T 2 relaxation time. TDM provided similar diffusion and relaxation parameters as SE using REDIM, whereas SB‐ME exhibited a 60% larger bias in the map and on average 3.5× larger bias in the covariance between relaxation‐diffusion coefficients.
Conclusion
Our analysis demonstrates that TDM provides a more accurate estimation of relaxation‐diffusion measurements while accelerating the acquisitions by a factor of 2 to 3.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Brigham and Women's Hospital, Harvard Medical School, Boston, Massachusetts, USA.
- School of Biomedical Engineering, Southern Medical University, Guangzhou, China.
- Department of Radiology, Harvard Medical School, Boston, Massachusetts, USA.
- Fetal-Neonatal Neuroimaging & Developmental Science Center, Boston Children's Hospital, Boston, Massachusetts, USA.
- Va Salt Lake City Health Care System, Informatics, Decision-Enhancement and Analytic Sciences Center, Salt Lake City, Utah, USA.
- Department of Neurology, University of Utah, Salt Lake City, Utah, USA.
ORCID-Profile: Qiang Liu, Borjan Gagoski, Berkin Bilgic
Förderung
- NIMH NIH HHS
- NINDS NIH HHS
- NIBIB NIH HHS
- U.S. Department of Veterans Affairs
- National Institute of Mental Health
- National Institute of Biomedical Imaging and Bioengineering
- National Institute of Neurological Disorders and Stroke
Zitieren
APA
Liu, Q., Gagoski, B., Shaik, I. A., Westin, C. F., Wilde, E. A., Schneider, W., Bilgic, B., Grissom, W. A., Nielsen, J. F., Zaitsev, M., Rathi, Y., & Ning, L. (2024). Time‐division multiplexing ( TDM ) sequence removes bias in T 2 estimation and relaxation‐diffusion measurements. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.30246
Vancouver
Liu Q, Gagoski B, Shaik IA, Westin CF, Wilde EA, Schneider W, et al. Time‐division multiplexing ( TDM ) sequence removes bias in T 2 estimation and relaxation‐diffusion measurements. Magnetic Resonance in Medicine. 2024. doi:10.1002/mrm.30246
BibTeX
@article{Liu2024,
author = {Liu, Qiang and Gagoski, Borjan and Shaik, Imam Ahmed and Westin, Carl-Fredrik and Wilde, Elisabeth A and Schneider, Walter and Bilgic, Berkin and Grissom, William A and Nielsen, Jon-Fredrik and Zaitsev, Maxim and Rathi, Yogesh and Ning, Lipeng},
title = {Time‐division multiplexing ( TDM ) sequence removes bias in T 2 estimation and relaxation‐diffusion measurements},
journal = {Magnetic Resonance in Medicine},
year = {2024},
doi = {10.1002/mrm.30246}
}
EUVIMED