Characterization of kilovoltage x‐ray image guidance system with a novel post‐processing algorithm on a new slip ring‐mounted radiotherapy system
Journal of Applied Clinical Medical Physics · 2024
DOI: 10.1002/acm2.14524 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2024. Verfasst von Heling Zhu und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, Peking Union Medical College Hospital, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijin…. Hauptthemen laut MEDLINE: Algorithms, Phantoms, Imaging, Cone-Beam Computed Tomography, Particle Accelerators. Die Arbeit nennt 21 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: National Key R&D Program of China, Ministry of Science and Technology of the People's Republic of China, National High Level Hospital Clinical Research Funding.
Abstract (Originalsprache)
Purpose
This study evaluates the performance of a kilovoltage x‐ray image‐guidance system equipped with a novel post‐processing optimization algorithm on the newly introduced TAICHI linear accelerator (Linac).
Methods
A comparative study involving image quality tests and radiation dose measurements was conducted across six scanning protocols of the kV‐cone beam computed tomography (CBCT) system on the TAICHI Linac. The performance assessment utilized the conventional Feldkamp–Davis–Kress (FDK) algorithm and a novel Non‐Local Means denoising and adaptive scattering correction (NLM‐ASC) algorithm. Image quality metrics, including spatial resolution, contrast‐to‐noise ratio (CNR), and signal‐to‐noise ratio (SNR), were evaluated using a Catphan 604 phantom. Radiation doses for low‐dose and standard protocols were measured using a computed tomography dose index (CTDI) phantom, with comparative measurements from the Halcyon Linac's iterative CBCT (iCBCT).
Results
The NLM‐ASC algorithm significantly improved image quality, achieving a 300%–1000% increase in CNR and SNR over the FDK‐only images and it also showed a 100%–200% improvement over the iCBCT images from Halcyon's head protocol. The optimized low‐dose protocols yielded higher image quality than the standard FDK protocols, indicating potential for reduced radiation exposure. Clinical implementation confirmed the TAICHI system's utility for precise and adaptive radiotherapy.
Conclusion
The kV‐IGRT system on the TAICHI Linac, with its novel post‐processing algorithm, demonstrated superior image quality suitable for routine clinical use, effectively reducing image noise without compromising other quality metrics.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiation Oncology, Peking Union Medical College Hospital, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing, China.
- Our United Corporation, Xi'an, Shanxi Province, China.
ORCID-Profile: Heling Zhu
Förderung
- National Key R&D Program of China
- Ministry of Science and Technology of the People's Republic of China
- National High Level Hospital Clinical Research Funding
- National Key Research and Development Program of China
Zitieren
APA
Zhu, H., Dong, T., Pang, T., Guan, Q., Yang, J., Zhao, F., Yang, B., & Qiu, J. (2024). Characterization of kilovoltage x‐ray image guidance system with a novel post‐processing algorithm on a new slip ring‐mounted radiotherapy system. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.14524
Vancouver
Zhu H, Dong T, Pang T, Guan Q, Yang J, Zhao F, et al. Characterization of kilovoltage x‐ray image guidance system with a novel post‐processing algorithm on a new slip ring‐mounted radiotherapy system. Journal of Applied Clinical Medical Physics. 2024. doi:10.1002/acm2.14524
BibTeX
@article{Zhu2024,
author = {Zhu, Heling and Dong, Tingting and Pang, Tingtian and Guan, Qiu and Yang, Jingru and Zhao, Feini and Yang, Bo and Qiu, Jie},
title = {Characterization of kilovoltage x‐ray image guidance system with a novel post‐processing algorithm on a new slip ring‐mounted radiotherapy system},
journal = {Journal of Applied Clinical Medical Physics},
year = {2024},
doi = {10.1002/acm2.14524}
}
EUVIMED