Image quality characterization of an ultra‐high‐speed kilovoltage cone‐beam computed tomography imaging system on an O‐ring linear accelerator

Euidam Kim iD, Yang Kyun Park iD, Tianyu Zhao, Eric Laugeman, Xiaodong Neo Zhao iD, Yao Hao iD, Yoonsun Chung, Hugh Lee iD

Journal of Applied Clinical Medical Physics · 2024-05 · 25 (5)

DOI: 10.1002/acm2.14337 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, online erschienen am 04.04.2024. Verfasst von Euidam Kim und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, Washington University in St Louis School of Medicine, St Louis, Missouri, USA. Hauptthemen laut MEDLINE: Cone-Beam Computed Tomography, Particle Accelerators, Phantoms, Imaging, Radiotherapy Planning, Computer-Assisted. Die Arbeit wurde bisher 56-mal zitiert (Crossref) und nennt 22 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

The quality of on‐board imaging systems, including cone‐beam computed tomography (CBCT), plays a vital role in image‐guided radiation therapy (IGRT) and adaptive radiotherapy. Recently, there has been an upgrade of the CBCT systems fused in the O‐ring linear accelerators called HyperSight, featuring a high imaging performance. As the characterization of a new imaging system is essential, we evaluated the image quality of the HyperSight system by comparing it with Halcyon 3.0 CBCT and providing benchmark data for routine imaging quality assurance.

Methods

The HyperSight features ultra‐fast scan time, a larger kilovoltage (kV) detector, a more substantial kV tube, and an advanced reconstruction algorithm. Imaging protocols in the two modes of operation, treatment mode with IGRT and the CBCT for planning (CBCTp) mode were evaluated and compared with Halcyon 3.0 CBCT. Image quality metrics, including spatial resolution, contrast resolution, uniformity, noise, computed tomography (CT) number linearity, and calibration error, were assessed using a Catphan and an electron density phantom and analyzed with TotalQA software.

Results

HyperSight demonstrated substantial improvements in contrast‐to‐noise ratio and noise in both IGRT and CBCTp modes compared to Halcyon 3.0 CBCT. CT number calibration error of HyperSight CBCTp mode (1.06%) closely matches that of a full CT scanner (0.72%), making it suitable for adaptive planning. In addition, the advanced hardware of HyperSight, such as ultra‐fast scan time (5.9 s) or 2.5 times larger heat unit capacity, enhanced the clinical efficiency in our experience.

Conclusions

HyperSight represented a significant advancement in CBCT imaging. With its image quality, CT number accuracy, and ultra‐fast scans, HyperSight has a potential to transform patient care and treatment outcomes. The enhanced scan speed and image quality of HyperSight are expected to significantly improve the quality and efficiency of treatment, particularly benefiting patients.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Cone-Beam Computed TomographyParticle AcceleratorsPhantoms, ImagingRadiotherapy Planning, Computer-AssistedImage Processing, Computer-AssistedRadiotherapy, Image-GuidedAlgorithmsRadiotherapy DosageHumansRadiotherapy, Intensity-ModulatedQuality Assurance, Health CareRadiographic Image Interpretation, Computer-Assisted

Stichwörter der Autor:innen

cone‐beam computational tomography (CBCT)image characterizationimage qualityimaging quality assurance (QA)

Autor:innen und Institutionen

  • Department of Radiation Oncology, Washington University in St Louis School of Medicine, St Louis, Missouri, USA.
  • Department of Nuclear Engineering, Hanyang University College of Engineering, Seoul, South Korea.
  • Department of Radiation Oncology, University of Texas Southwestern Medical Center, Dallas, Texas, USA.

ORCID-Profile: Euidam Kim, Yang Kyun Park, Xiaodong Neo Zhao, Yao Hao, Hugh Lee

Zitieren

APA

Kim, E., Park, Y. K., Zhao, T., Laugeman, E., Zhao, X. N., Hao, Y., Chung, Y., & Lee, H. (2024). Image quality characterization of an ultra‐high‐speed kilovoltage cone‐beam computed tomography imaging system on an O‐ring linear accelerator. Journal of Applied Clinical Medical Physics, 25(5). https://doi.org/10.1002/acm2.14337

Vancouver

Kim E, Park YK, Zhao T, Laugeman E, Zhao XN, Hao Y, et al. Image quality characterization of an ultra‐high‐speed kilovoltage cone‐beam computed tomography imaging system on an O‐ring linear accelerator. Journal of Applied Clinical Medical Physics. 2024;25(5). doi:10.1002/acm2.14337

BibTeX
@article{Kim2024,
  author = {Kim, Euidam and Park, Yang Kyun and Zhao, Tianyu and Laugeman, Eric and Zhao, Xiaodong Neo and Hao, Yao and Chung, Yoonsun and Lee, Hugh},
  title = {Image quality characterization of an ultra‐high‐speed kilovoltage cone‐beam computed tomography imaging system on an O‐ring linear accelerator},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2024},
  volume = {25},
  number = {5},
  doi = {10.1002/acm2.14337}
}
Kostenlos starten →