HyperSight CBCT image quality and metal artifact reduction for adaptive head and neck radiotherapy: Results from a prospective clinical trial

Abby Yashayaeva, R Lee MacDonald, Kenny Zhan, Jennifer DeGiobbi, Natasha McMaster, Dave McAloney, Lucy Ward, Cheryl Anderson, Marc LeBlanc, Lara Best, Murali Rajaraman, Derek Wilke und 1 weitere

Journal of Applied Clinical Medical Physics · 2025-12 · 26 (12)

DOI: 10.1002/acm2.70338 ↗

Kurz gefasst

Klinische Studie in der Zeitschrift Journal of Applied Clinical Medical Physics, online erschienen am 29.11.2025. Verfasst von Abby Yashayaeva und 12 weiteren Autor:innen. Beteiligte Einrichtung: Department of Physics and Atmospheric Sciences, Dalhousie University, Halifax, Nova Scotia, Canada. Hauptthemen laut MEDLINE: Artifacts, Cone-Beam Computed Tomography, Head and Neck Neoplasms, Image Processing, Computer-Assisted. Die Arbeit wurde bisher 10-mal zitiert (Crossref) und nennt 43 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Accurate Hounsfield units (HU) are critical for dose calculation and anatomical visualization, but are often affected by dental artifacts in head and neck (H&N) cancer patients. The HyperSight cone‐beam computed tomography (CBCT) platform provides improved image quality over previous CBCT platforms and offers metal artifact reduction (iCBCT MAR) reconstruction.

Purpose

This study evaluates the quality of HyperSight CBCT images compared to current clinical standards: TrueBeam CBCT for image guidance and fan‐beam CT (FBCT) from a CT simulator for treatment planning, using images captured during H&N cancer treatment.

Methods

Images for 30 H&N cancer patients were acquired on a HyperSight CBCT, conventional TrueBeam and FBCT, with 24 patients exhibiting metal dental artifacts. The HyperSight images were reconstructed using iCBCT MAR and iterative (iCBCT Acuros) algorithms. The four image sets were rigidly registered and compared using the artifact index (AI) measured in the oral cavity and the percentile range (PR) measured in the oral cavity, brain, brainstem and eyes to assess image non‐uniformity. The HU accuracy was calculated relative to FBCT (baseline) for soft tissues (oral cavity, brainstem, submandibular and parotid glands), and bone (mandible). The contrast relative to baseline was evaluated between the oral cavity and nearby structures. Image‐based metrics were computed relative to FBCT including structural similarity index measure (SSIM), mean‐square error (MSE) and peak signal‐to‐noise ratio (PSNR).

Results

The HyperSight iCBCT MAR images showed a significant reduction in AI values compared to the other images ( p < 0.0004), but higher PR values indicating decreased HU uniformity compared to HyperSight iCBCT Acuros and FBCT ( p < 0.0002). The soft‐tissue HU and contrast values were significantly closer to baseline in both HyperSight images compared to TrueBeam (HU: p < 0.001, contrast: p < 0.001). For soft‐tissue the HU mean absolute deviation (MAD) from baseline was 16 ± 10 HU for HyperSight iCBCT Acuros, 15 ± 10 HU for HyperSight iCBCT MAR, and 35 ± 22 HU for TrueBeam. For bone, the HU MAD from baseline was 153 ± 233 HU, 185 ± 268 HU, and 214 ± 212 HU, respectively. The HyperSight iCBCT Acuros algorithm achieved significantly superior SSIM, MSE, and PSNR metrics compared to TrueBeam and HyperSight iCBCT MAR in regions with large amounts of bone and air.

Conclusions

HyperSight iCBCT MAR significantly reduced artifacts compared to HyperSight iCBCT Acuros, TrueBeam and FBCT, making it particularly beneficial for patients with metal implants. Both HyperSight reconstructions demonstrated improved soft‐tissue HU accuracy and contrast compared to TrueBeam, however the iCBCT Acuros algorithm may be preferred when metal‐induced artifacts are not a concern. These results support the suitability of HyperSight images in adaptive treatment workflows requiring accurate image quality, even with severe metal artifacts.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

ArtifactsCone-Beam Computed TomographyHead and Neck NeoplasmsImage Processing, Computer-AssistedMetalsRadiotherapy Planning, Computer-AssistedRadiotherapy, Image-GuidedFemaleHumansMaleAlgorithmsPhantoms, ImagingProspective StudiesRadiotherapy DosageRadiotherapy, Intensity-Modulated

Stichwörter der Autor:innen

adaptive radiotherapycone beam computed tomographymetal artifact reduction

Substanzen

Metals

Autor:innen und Institutionen

  • Department of Physics and Atmospheric Sciences, Dalhousie University, Halifax, Nova Scotia, Canada.
  • Department of Radiation Oncology, QE2 Cancer Centre, Nova Scotia Health, Halifax, Nova Scotia, Canada.
  • Department of Radiation Oncology, Dalhousie University, Halifax, Nova Scotia, Canada.

Zitieren

APA

Yashayaeva, A., MacDonald, R. L., Zhan, K., DeGiobbi, J., McMaster, N., McAloney, D., Ward, L., Anderson, C., LeBlanc, M., Best, L., Rajaraman, M., Wilke, D., & Cherpak, A. (2025). HyperSight CBCT image quality and metal artifact reduction for adaptive head and neck radiotherapy: Results from a prospective clinical trial. Journal of Applied Clinical Medical Physics, 26(12). https://doi.org/10.1002/acm2.70338

Vancouver

Yashayaeva A, MacDonald RL, Zhan K, DeGiobbi J, McMaster N, McAloney D, et al. HyperSight CBCT image quality and metal artifact reduction for adaptive head and neck radiotherapy: Results from a prospective clinical trial. Journal of Applied Clinical Medical Physics. 2025;26(12). doi:10.1002/acm2.70338

BibTeX
@article{Yashayaeva2025,
  author = {Yashayaeva, Abby and MacDonald, R Lee and Zhan, Kenny and DeGiobbi, Jennifer and McMaster, Natasha and McAloney, Dave and Ward, Lucy and Anderson, Cheryl and LeBlanc, Marc and Best, Lara and Rajaraman, Murali and Wilke, Derek and Cherpak, Amanda},
  title = {HyperSight CBCT image quality and metal artifact reduction for adaptive head and neck radiotherapy: Results from a prospective clinical trial},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2025},
  volume = {26},
  number = {12},
  doi = {10.1002/acm2.70338}
}
Kostenlos starten →