EUVIMEDEUVIMED entdecken
FallberichtOpen Access

Sub‐Acute Re‐Elevated Pacing Thresholds After Atrial Leadless Pacemaker Implantation

Dongyan Wu iD, Le He, Fengmin Lu, Ying Shao, Yina Li, Enyuan Zhang iD, Henan Zhang, Jing Xu

Journal of Cardiovascular Electrophysiology · 2026

DOI: 10.1111/jce.70478 ↗

Kurz gefasst

Fallbericht in der Zeitschrift Journal of Cardiovascular Electrophysiology, 2026. Verfasst von Dongyan Wu und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cardiology, Chest Hospital, Tianjin University, Tianjin, China. Hauptthemen laut MEDLINE: Pacemaker, Artificial, Cardiac Pacing, Artificial, Bradycardia, Heart Rate. Die Arbeit nennt 7 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: Tianjin Key Medical Discipline Construction Project.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Introduction

The safety and efficacy of the atrial leadless pacemaker (A‐LP) with helix fixation have been demonstrated. Usually, elevated acute thresholds decrease gradually and normalize within 24 h. Here, we report a case of subacute re‐elevation of pacing thresholds after hospital discharge. Case Presentation An 80‐year‐old female with a history of bradycardia underwent A‐LP implantation. At the time of the procedure, the final pacing threshold (PT) was measured at 4.0 V/0.4 ms. On post‐procedural Day 1, the PT at 0.4 ms pulse width decreased to 1.0 V. On Day 4, the PT was 1.0 V/0.4 ms prior to the patient's discharge. However, on Day 5, she was readmitted to the hospital due to the loss of atrial capture at the maximum output available after an intravenous administration of 150 mg amiodarone for atrial flutter. Over 3 days of monitoring, atrial PT was higher than 6 V/1.5 ms. A traditional DDD pacemaker was implanted due to recurrent sinus asystole associated with dizziness. At the 1‐month follow‐up, the PT of atrial A‐LP was 0.5 V/0.4 ms. Two months later, an ambulatory Holter confirmed the stable and normal atrial capture of the A‐LP after lowering the base rate of the DDD pacemaker. At the 3‐month follow‐up, PT of the A‐LP remained 0.5 V/0.4 ms.

Conclusion

Sub‐acute re‐elevated PT of the A‐LP during the perioperative period may persist for more than 1 week and then normalize within 1 month. If PT re‐elevation is suspected to be related to peri‐electrode inflammation, a conservative approach with continued monitoring of threshold changes may be considered.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Pacemaker, ArtificialCardiac Pacing, ArtificialBradycardiaHeart RateAtrial FlutterHumansFemaleAged, 80 and overTreatment OutcomeTime FactorsAction Potentials

Stichwörter der Autor:innen

amiodaroneatrial leadless pacemakerpacing threshold elevation

Autor:innen und Institutionen

  • Department of Cardiology, Chest Hospital, Tianjin University, Tianjin, China.

ORCID-Profile: Dongyan Wu, Enyuan Zhang

Förderung

  • Tianjin Key Medical Discipline Construction Project

Zitieren

APA

Wu, D., He, L., Lu, F., Shao, Y., Li, Y., Zhang, E., Zhang, H., & Xu, J. (2026). Sub‐Acute Re‐Elevated Pacing Thresholds After Atrial Leadless Pacemaker Implantation. Journal of Cardiovascular Electrophysiology. https://doi.org/10.1111/jce.70478

Vancouver

Wu D, He L, Lu F, Shao Y, Li Y, Zhang E, et al. Sub‐Acute Re‐Elevated Pacing Thresholds After Atrial Leadless Pacemaker Implantation. Journal of Cardiovascular Electrophysiology. 2026. doi:10.1111/jce.70478

BibTeX
@article{Wu2026,
  author = {Wu, Dongyan and He, Le and Lu, Fengmin and Shao, Ying and Li, Yina and Zhang, Enyuan and Zhang, Henan and Xu, Jing},
  title = {Sub‐Acute Re‐Elevated Pacing Thresholds After Atrial Leadless Pacemaker Implantation},
  journal = {Journal of Cardiovascular Electrophysiology},
  year = {2026},
  doi = {10.1111/jce.70478}
}
Kostenlos starten →