Comparison of outcomes and required tools between transvenous extraction of pacemaker and implantable cardioverter defibrillator leads: Insight from single high‐volume center experience
Journal of Cardiovascular Electrophysiology · 2024-07 · 35 (7) · 1382-1392
DOI: 10.1111/jce.16294 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift Journal of Cardiovascular Electrophysiology, online erschienen am 09.05.2024. Verfasst von Katsuhide Hayashi und 8 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cardiovascular Medicine, Cardiac Electrophysiology and Pacing Section, Cleveland Clinic, Cleveland, Ohio, USA. Hauptthemen laut MEDLINE: Defibrillators, Implantable, Pacemaker, Artificial, Device Removal, Registries. Die Arbeit wurde bisher 8-mal zitiert (Crossref) und nennt 16 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).
Abstract (Originalsprache)
Introduction
Reports of comparison with procedural outcomes for implantable cardioverter defibrillator (ICD) and pacemaker (PM) transvenous lead extraction (TLE) are old and limited. We sought to compare the safety, efficacy, and procedural properties of ICD and PM TLE and assess the impact of lead age.
Methods
The study cohort included all consecutive patients with ICD and PM TLE in the Cleveland Clinic Prospective TLE Registry between 2013 and 2022. Extraction success, complications, and failure employed the definitions described in the HRS 2017 TLE guidelines.
Results
A total of 885 ICD leads, a median implant duration of 8 (5−11) years in 810 patients, and 1352 PM leads of 7 (3−13) years in 807 patients were included. Procedural success rates in ICD patients were superior to those of PM in >20 years leads but similar in ≤20 years leads. In the PM group, the complete success rate of TLE decreased significantly according to the increase of lead age, but not in the ICD group. ICD TLE required more extraction tools compared with PM TLE but cases with older leads required non‐laser sheath extraction tools in both groups. The most common injury site in major complication cases differed between ICD and PM TLE, although major complication rates showed no difference in both groups (2.7% vs. 1.6%, p = .12).
Conclusion
The procedural success rate by TLE is greater for ICD patients than PM patients with leads >20 years old but requires more extraction tools. Common vascular complication sites and the impact of lead age on procedural outcomes and required tools differed between ICD and PM TLE.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Cardiovascular Medicine, Cardiac Electrophysiology and Pacing Section, Cleveland Clinic, Cleveland, Ohio, USA.
ORCID-Profile: Katsuhide Hayashi, Thomas Callahan, John Rickard, Shady Nakhla, Chadi Tabaja, Aritra Paul
Zitieren
APA
Hayashi, K., Callahan, T., Rickard, J., Baranowski, B., Martin, D. O., Nakhla, S., Tabaja, C., Paul, A., & Wilkoff, B. L. (2024). Comparison of outcomes and required tools between transvenous extraction of pacemaker and implantable cardioverter defibrillator leads: Insight from single high‐volume center experience. Journal of Cardiovascular Electrophysiology, 35(7), 1382-1392. https://doi.org/10.1111/jce.16294
Vancouver
Hayashi K, Callahan T, Rickard J, Baranowski B, Martin DO, Nakhla S, et al. Comparison of outcomes and required tools between transvenous extraction of pacemaker and implantable cardioverter defibrillator leads: Insight from single high‐volume center experience. Journal of Cardiovascular Electrophysiology. 2024;35(7):1382-1392. doi:10.1111/jce.16294
BibTeX
@article{Hayashi2024,
author = {Hayashi, Katsuhide and Callahan, Thomas and Rickard, John and Baranowski, Bryan and Martin, David O and Nakhla, Shady and Tabaja, Chadi and Paul, Aritra and Wilkoff, Bruce L},
title = {Comparison of outcomes and required tools between transvenous extraction of pacemaker and implantable cardioverter defibrillator leads: Insight from single high‐volume center experience},
journal = {Journal of Cardiovascular Electrophysiology},
year = {2024},
volume = {35},
number = {7},
pages = {1382--1392},
doi = {10.1111/jce.16294}
}
EUVIMED