EUVIMEDEUVIMED entdecken
Vergleichsstudie

Removal of leads broken during extraction: A comparison of different approaches and tools

Andrzej Kutarski, Wojciech Jacheć, Radosław Pietura, Paweł Stefańczyk iD, Jarosław Kosior, Marek Czakowski iD, Sebastian Sawonik, Łukasz Tułecki, Dorota Nowosielecka iD

Journal of Cardiovascular Electrophysiology · 2024

DOI: 10.1111/jce.16398 ↗

Kurz gefasst

Vergleichsstudie in der Zeitschrift Journal of Cardiovascular Electrophysiology, 2024. Verfasst von Andrzej Kutarski und 8 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cardiology, Medical University of Lublin, Lublin, Poland. Hauptthemen laut MEDLINE: Device Removal, Equipment Failure, Pacemaker, Artificial, Defibrillators, Implantable. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 48 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Extraction of a broken lead fragment (BLF) has received scant attention in the literature.

Methods

Retrospective analysis was to compare the effectiveness of different approaches and tools used for BLF removal during 127 procedures.

Results

A superior approach was the most popular (75.6%), femoral (15.7%) and combined (8.7%) approaches were the least common. Of 127 BLFs 78 (61.4%) were removed in their entirety and BLF length was significantly reduced to less than 4 cm in 21 (16.5%) or lead tip in 12 (9.4%) cases. The best results were achieved when BLFs were longer (>4 cm) (62/93 66.7% of longer BLFs), either in the case of BLFs free‐floating in vascular bed including pulmonary circulation (68.4% of them) but not in cases of short BLFs (20.0% of short BLFs). Complete procedural success was achieved in 57.5% of procedures, the lead tip retained in the heart wall in 12 cases (9.4%) and short BLFs were found in 26.0%, whereas BLFs >4 cm were left in place in four cases (3.1%) of procedures only. There was no relationship between approach in lead remnant removal and long‐term mortality.

Conclusions

(1) Effectiveness of fractured lead removal is satisfactory: entire BLFs were removed in 61.4% (total procedural success—57.5%, was lower because five major complications occurred) and BLF length was significantly reduced in 26.0%. (2) Among the broken leads, leads with a long stay in the patient (16.3 years on average), passive leads (97.6%) and pacemaker leads 92.1% are significantly more common, but not ICD leads (only 7.9% of lead fractures) compared to TLE without lead fractures. (3) Broken lead removal (superior approach) using a CS access sheath as a “subclavian workstation” for continuation of dilatation with conventional tools deserves attention. (4) Lead fracture management should become an integral part of training in transvenous lead extraction.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Device RemovalEquipment FailurePacemaker, ArtificialDefibrillators, ImplantableHumansRetrospective StudiesFemaleMaleAgedMiddle AgedTreatment OutcomeAged, 80 and overAdult

Stichwörter der Autor:innen

femoral approachincomplete lead extractionlead fragment removaltechniques of lead fragment removaltransvenous lead extraction

Autor:innen und Institutionen

  • Department of Cardiology, Medical University of Lublin, Lublin, Poland.
  • 2nd Department of Cardiology, Faculty of Medical Sciences in Zabrze, Medical University of Silesia, Katowice, Poland.
  • Department of Radiography, Medical University of Lublin, Lublin, Poland.
  • Department of Cardiology, The Pope John Paul II Province Hospital, Zamość, Poland.
  • Department of Cardiology, Masovian Specialistic Hospital of Radom, Radom, Poland.
  • Department of Cardiac Surgery of Medical University, Medical University of Lublin, Lublin, Poland.

ORCID-Profile: Paweł Stefańczyk, Marek Czakowski, Dorota Nowosielecka

Zitieren

APA

Kutarski, A., Jacheć, W., Pietura, R., Stefańczyk, P., Kosior, J., Czakowski, M., Sawonik, S., Tułecki, Ł., & Nowosielecka, D. (2024). Removal of leads broken during extraction: A comparison of different approaches and tools. Journal of Cardiovascular Electrophysiology. https://doi.org/10.1111/jce.16398

Vancouver

Kutarski A, Jacheć W, Pietura R, Stefańczyk P, Kosior J, Czakowski M, et al. Removal of leads broken during extraction: A comparison of different approaches and tools. Journal of Cardiovascular Electrophysiology. 2024. doi:10.1111/jce.16398

BibTeX
@article{Kutarski2024,
  author = {Kutarski, Andrzej and Jacheć, Wojciech and Pietura, Radosław and Stefańczyk, Paweł and Kosior, Jarosław and Czakowski, Marek and Sawonik, Sebastian and Tułecki, Łukasz and Nowosielecka, Dorota},
  title = {Removal of leads broken during extraction: A comparison of different approaches and tools},
  journal = {Journal of Cardiovascular Electrophysiology},
  year = {2024},
  doi = {10.1111/jce.16398}
}
Kostenlos starten →