The Test‐Retest Reliability and Minimal Detectable Change of the Brain Pro System for Assessing Upper Extremity Visuo‐Motor Reaction Time in Physically Active Healthy Individuals
Physiotherapy Research International · 2026-04 · 31 (2)
DOI: 10.1002/pri.70171 ↗Astratt
ABSTRACT Background and Purpose To determine the test‐retest reliability and minimal detectable change of the Brain Pro System (Brain Pro System, Poland) for assessing upper extremity visuo‐motor reaction time (V‐MRT) in physically active healthy individuals. Methods Sixty‐four physically active healthy participants (M age = 21.26, SD = 0.94 years) completed two assessments 1 week apart. Reliability was assessed using ICC2,1 with a 95% confidence interval. The standard error of measurement and MDC95 were calculated. Stability was analyzed with Cronbach's α and coefficient of variation (CV). Results The Brain Pro System demonstrated excellent reliability for right, left, and randomly determined upper extremity V‐MRT (ICC2,1 = 0.95–0.97, α = 0.95–0.97, CV = 0.01). The MDC values were 1.21 s for right and left V‐MRT and 1.10 s for combined extremities. Discussion The Brain Pro System is highly reliable for assessing V‐MRT. The MDC values provide a reference for evaluating meaningful changes in reaction time for physically active, healthy adults.