Free‐Breathing Fat Quantification Using a Phase Error‐Corrected Cartesian Acquisition With Spiral Profile Ordering

Philipp Braun iD, Jonathan Stelter, Stefan Ruschke iD, Johannes M Peeters, Holger Eggers, Kilian Weiss, Daniela Junker, Dimitrios C Karampinos

Magnetic Resonance in Medicine · 2026

DOI: 10.1002/mrm.70474 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2026. Verfasst von Philipp Braun und 7 weiteren Autor:innen. Beteiligte Einrichtung: Institute for Diagnostic and Interventional Radiology, School of Medicine and Health, TUM University Hospital, Technical University of Munich (TUM), …. Hauptthemen laut MEDLINE: Adipose Tissue, Magnetic Resonance Imaging, Image Processing, Computer-Assisted. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 63 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Deutsche Forschungsgemeinschaft.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

To develop a phase‐corrected time‐interleaved multi‐echo gradient echo Cartesian acquisition with spiral profile ordering (TIMGRECASPR) for abdominal large‐FOV proton density fat fraction (PDFF) mapping at 3 T , demonstrating its sampling flexibility and inherent self‐gating capabilities at high isotropic resolutions.

Methods

A methodological framework was developed to correct eddy current‐ and concomitant gradient‐related phase errors in a TIMGRECASPR acquisition. Sequence parameters were optimized with an XCAT simulation. Validation was conducted in phantoms and in vivo thigh and abdominal imaging. In five healthy volunteers, liver PDFF and T2 *, alongside pancreatic PDFF, were compared to a clinical Cartesian breath‐hold vendor technique. Liver PDFF homogeneity was evaluated visually and across five different axial subvolumes. Mean absolute PDFF bias and within‐subject standard deviation (wSD) across repeated scans were quantified. Pancreatic delineation was quantified by line profiles across the tissue boundary.

Results

Short shot lengths were required for a sufficient temporal resolution of the respiratory motion self‐navigation. Without phase corrections, data showed severe peripheral PDFF bias, especially at higher spatial resolutions. In a water phantom, PDFF bias ranged from −5% to 5% over 400 mm in readout direction, successfully mitigated by the proposed phase corrections. Liver PDFF maps exhibited improved homogeneity and accurate quantification results, verified by the breath‐hold reference scan (0.6% PDFF bias, wSD = 0.11%). Pancreatic boundaries were clearly visualized and provided accurate pancreatic PDFF maps (1.1% PDFF bias, wSD = 0.37%).

Conclusion

This work demonstrates the feasibility of a free‐breathing monopolar time‐interleaved gradient echo sequence based on a Cartesian spiral acquisition at 3 T when accounting for phase errors.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Adipose TissueMagnetic Resonance ImagingImage Processing, Computer-AssistedHumansPhantoms, ImagingLiverAlgorithmsAbdomenBreath HoldingMaleRespirationPancreasAdultFemaleReproducibility of ResultsThighHealthy VolunteersProtons

Stichwörter der Autor:innen

Cartesian spiral profile ordering (CASPR)free‐breathingisotropic resolutionphase errorsproton density fat fraction (PDFF)time‐interleaved multi‐echo gradient echo

Substanzen

Protons

Autor:innen und Institutionen

  • Institute for Diagnostic and Interventional Radiology, School of Medicine and Health, TUM University Hospital, Technical University of Munich (TUM), Munich, Germany.
  • Philips Healthcare, Best, the Netherlands.
  • Philips GmbH Innovative Technologies, Hamburg, Germany.
  • Philips GmbH Market DACH, Hamburg, Germany.
  • Laboratory of Magnetic Resonance Imaging Systems and Methods, Ecole Polytechnique Federale de Lausanne (EPFL), Lausanne, Switzerland.
  • CIBM Center for Biomedical Imaging (CIBM), Lausanne, Switzerland.

ORCID-Profile: Philipp Braun, Stefan Ruschke

Förderung

  • Deutsche Forschungsgemeinschaft

Zitieren

APA

Braun, P., Stelter, J., Ruschke, S., Peeters, J. M., Eggers, H., Weiss, K., Junker, D., & Karampinos, D. C. (2026). Free‐Breathing Fat Quantification Using a Phase Error‐Corrected Cartesian Acquisition With Spiral Profile Ordering. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.70474

Vancouver

Braun P, Stelter J, Ruschke S, Peeters JM, Eggers H, Weiss K, et al. Free‐Breathing Fat Quantification Using a Phase Error‐Corrected Cartesian Acquisition With Spiral Profile Ordering. Magnetic Resonance in Medicine. 2026. doi:10.1002/mrm.70474

BibTeX
@article{Braun2026,
  author = {Braun, Philipp and Stelter, Jonathan and Ruschke, Stefan and Peeters, Johannes M and Eggers, Holger and Weiss, Kilian and Junker, Daniela and Karampinos, Dimitrios C},
  title = {Free‐Breathing Fat Quantification Using a Phase Error‐Corrected Cartesian Acquisition With Spiral Profile Ordering},
  journal = {Magnetic Resonance in Medicine},
  year = {2026},
  doi = {10.1002/mrm.70474}
}
Kostenlos starten →