Multishot Dual Polarity GRAPPA : Robust Nyquist Ghost Correction for Multishot EPI
Magnetic Resonance in Medicine · 2026-05 · 95 (5) · 2726-2736
DOI: 10.1002/mrm.70233 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, online erschienen am 25.12.2025. Verfasst von Yuancheng Jiang und 6 weiteren Autor:innen. Beteiligte Einrichtung: Center for Biomedical Imaging Research, School of Biomedical Engineering, Tsinghua University, Beijing, China. Hauptthemen laut MEDLINE: Echo-Planar Imaging, Image Processing, Computer-Assisted. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 55 Literaturangaben. Gefördert von: NIH HHS, National Institutes of Health.
Abstract (Originalsprache)
Purpose
This work aims to develop a robust Nyquist ghost correction method for multishot echo‐planar imaging (EPI). The method helps correct challenging Nyquist ghosts, particularly on scanners with high‐performance gradients or ultra‐high fields.
Methods
A method for multishot EPI ghost correction, called multishot dual‐polarity GRAPPA (msDPG), is developed by extending the DPG concept to multishot readouts. msDPG employs tailored DPG kernels to address high‐order phase differences between two EPI readout polarities, which cannot be fully addressed using linear phase correction (LPC) or non‐linear phase correction (nLPC). Advanced regularizers can be readily employed with the proposed msDPG for physiologic inter‐shot phase variation correction during reconstruction. Additionally, a calibration refinement method is proposed to improve the quality of the DPG calibration data and enhance reconstruction performance.
Results
Phantom and in vivo experiments on scanners with high‐performance gradients and ultra‐high fields demonstrated that msDPG achieved superior ghost correction performance than LPC and nLPC, reducing the ghost‐to‐signal ratio (GSR) by over 50%. Compared to conventional DPG, msDPG provided images with lower noise amplification, particularly for acquisitions with large in‐plane acceleration. Consequently, high‐fidelity, submillimeter diffusion images were obtained using msDPG with regularized reconstruction.
Conclusion
The proposed msDPG provides a robust Nyquist ghost correction method for multishot EPI, enabling submillimeter imaging with improved fidelity.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Center for Biomedical Imaging Research, School of Biomedical Engineering, Tsinghua University, Beijing, China.
- Athinoula A. Martinos Center for Biomedical Imaging, Massachusetts General Hospital, Charlestown, Massachusetts, USA.
- Harvard Medical School, Boston, Massachusetts, USA.
- Department of Psychiatry and Radiology, Brigham and Women's Hospital, Boston, Massachusetts, USA.
ORCID-Profile: Yohan Jun, Qiang Liu, Hua Guo, Berkin Bilgic
Förderung
- NIH HHS
- National Institutes of Health
Zitieren
APA
Jiang, Y., Jun, Y., Liu, Q., Zhong, W., Rathi, Y., Guo, H., & Bilgic, B. (2026). Multishot Dual Polarity GRAPPA : Robust Nyquist Ghost Correction for Multishot EPI. Magnetic Resonance in Medicine, 95(5), 2726-2736. https://doi.org/10.1002/mrm.70233
Vancouver
Jiang Y, Jun Y, Liu Q, Zhong W, Rathi Y, Guo H, et al. Multishot Dual Polarity GRAPPA : Robust Nyquist Ghost Correction for Multishot EPI. Magnetic Resonance in Medicine. 2026;95(5):2726-2736. doi:10.1002/mrm.70233
BibTeX
@article{Jiang2026,
author = {Jiang, Yuancheng and Jun, Yohan and Liu, Qiang and Zhong, Wen and Rathi, Yogesh and Guo, Hua and Bilgic, Berkin},
title = {Multishot Dual Polarity GRAPPA : Robust Nyquist Ghost Correction for Multishot EPI},
journal = {Magnetic Resonance in Medicine},
year = {2026},
volume = {95},
number = {5},
pages = {2726--2736},
doi = {10.1002/mrm.70233}
}
EUVIMED