Time‐resolved MR fingerprinting for T 2 * signal extraction: MR fingerprinting meets echo planar time‐resolved imaging
Magnetic Resonance in Medicine · 2025
DOI: 10.1002/mrm.30381 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2025. Verfasst von Di Cui und 3 weiteren Autor:innen. Beteiligte Einrichtung: Radiology & Biomedical Imaging, University of California San Francisco, San Francisco, California, USA. Hauptthemen laut MEDLINE: Echo-Planar Imaging, Brain, Image Processing, Computer-Assisted, Magnetic Resonance Imaging. Die Arbeit nennt 46 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: NIBIB NIH HHS, University of California, San Francisco, Department of Radiology & Biomedical Imaging, National Institute of Biomedical Imaging and Bioengineering.
Abstract (Originalsprache)
Purpose
This study leverages the echo planar time‐resolved imaging (EPTI) concept in MR fingerprinting (MRF) framework for a new time‐resolved MRF (TRMRF) approach, and explores its capability for fast simultaneous quantification of multiple MR parameters including T 1 , T 2 , T 2 *, proton density, off resonance, and B 1 + .
Methods
The proposed TRMRF method uses the concept of EPTI to track the signal change along the EPI echo train for T 2 * weighting with a k‐t Poisson‐based sampling order designed for acquisition. A two‐dimensional decomposition algorithm was designed for the image reconstruction, enabling fast and precise subspace modeling. The accuracy of proposed method was evaluated by a T 1 /T 2 phantom. The feasibility was demonstrated through 5 healthy volunteer brain studies.
Results
In the phantom studies, T 1 , T 2 , and T 2 * maps of TRMRF correlated strongly with gold‐standard methods. The concordance correlation coefficients are 0.9999, 0.9984 and 0.9978, and R 2 s are 0.9998, 0.9971, and 0.9983. In the in vivo studies, quantitative maps were acquired with 5 healthy volunteers. TRMRF was demonstrated to have comparable results with spiral MRF and gradient‐echo EPTI. TRMRF scans using 16, 10, and 6 s per slice were also evaluated to demonstrate the capability of shorter scan times.
Conclusion
A new approach is proposed to exploit the advantage of EPTI in the MRF framework. We demonstrate in phantom and in vivo experiments that T 1 , T 2 , T 2 *, proton density, off resonance, and B 1 + can be simultaneously quantified within 6 s/slice by TRMRF.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Radiology & Biomedical Imaging, University of California San Francisco, San Francisco, California, USA.
- Graduate Program in Bioengineering, University of California, Berkeley and San Francisco, San Francisco, California, USA.
ORCID-Profile: Di Cui, Xiaoxi Liu, Peder E Z Larson
Förderung
- NIBIB NIH HHS
- University of California, San Francisco, Department of Radiology & Biomedical Imaging
- National Institute of Biomedical Imaging and Bioengineering
Zitieren
APA
Cui, D., Liu, X., Larson, P. E. Z., & Xu, D. (2025). Time‐resolved MR fingerprinting for T 2 * signal extraction: MR fingerprinting meets echo planar time‐resolved imaging. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.30381
Vancouver
Cui D, Liu X, Larson PEZ, Xu D. Time‐resolved MR fingerprinting for T 2 * signal extraction: MR fingerprinting meets echo planar time‐resolved imaging. Magnetic Resonance in Medicine. 2025. doi:10.1002/mrm.30381
BibTeX
@article{Cui2025,
author = {Cui, Di and Liu, Xiaoxi and Larson, Peder E Z and Xu, Duan},
title = {Time‐resolved MR fingerprinting for T 2 * signal extraction: MR fingerprinting meets echo planar time‐resolved imaging},
journal = {Magnetic Resonance in Medicine},
year = {2025},
doi = {10.1002/mrm.30381}
}