Design and operation of a whole‐body MRI scanner without RF shielding

Stephan Biber, Stephan Kannengiesser, Juergen Nistler, Markus Braun, Stefan Blaess, Matthias Gebhardt, David Grodzki, Dieter Ritter, Georg Seegerer, Markus Vester, Rainer Schneider iD

Magnetic Resonance in Medicine · 2025-04 · 93 (4) · 1842-1855

DOI: 10.1002/mrm.30374 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, online erschienen am 11.11.2024. Verfasst von Stephan Biber und 10 weiteren Autor:innen. Beteiligte Einrichtung: Siemens Healthineers AG, Erlangen, Germany. Hauptthemen laut MEDLINE: Magnetic Resonance Imaging, Whole Body Imaging. Die Arbeit wurde bisher 11-mal zitiert (Crossref) und nennt 21 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

To investigate how a clinical MRI system can be operated without an RF cabin. Receive interference cancellation via signal processing and active transmit emission suppression are evaluated for mitigating image artifacts and achieving electromagnetic compatibility.

Methods

A clinical whole‐body MR scanner with B 0 = 0.55 T was placed in an environment without RF shielding. The signals of 16 additional auxiliary receive antenna elements outside the scanner bore were used to remove interference from the wanted MR signal via a dedicated sidelobe cancellation algorithm. In vivo measurements with different MR sequences were performed in the presence of ambient RF noise and additional artificial interferers, covering brain and abdominal applications. To investigate the transmit electromagnetic compatibility, simulations with different auxiliary transmit antenna arrangements and body coil loading conditions were performed, and their emission suppression performance in the far field was evaluated.

Results

The MR sidelobe cancellation algorithm successfully removed interference artifacts up to more than 3 times of the wanted MR image energy; difference images carried little to no visibly removed wanted signal. According to the simulations, electromagnetic compliance can be achieved with 32 or more auxiliary transmit antennas. The number and placement of sensor points are essential.

Conclusions

A whole‐body 0.55T MR scanner can be operated without RF shielding. Image artifacts from external interference can be removed, requiring only a modest number of auxiliary antennas. Electromagnetic compliance appears to be achievable using active transmit cancellation, but the required hardware efforts may not be practical.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Magnetic Resonance ImagingWhole Body ImagingHumansAlgorithmsArtifactsEquipment DesignPhantoms, ImagingRadio WavesComputer SimulationImage Processing, Computer-AssistedBrainSignal Processing, Computer-Assisted

Stichwörter der Autor:innen

RF shieldingaccessible MRInoise cancellationwhole‐body MRI

Autor:innen und Institutionen

  • Siemens Healthineers AG, Erlangen, Germany.

ORCID-Profile: Rainer Schneider

Zitieren

APA

Biber, S., Kannengiesser, S., Nistler, J., Braun, M., Blaess, S., Gebhardt, M., Grodzki, D., Ritter, D., Seegerer, G., Vester, M., & Schneider, R. (2025). Design and operation of a whole‐body MRI scanner without RF shielding. Magnetic Resonance in Medicine, 93(4), 1842-1855. https://doi.org/10.1002/mrm.30374

Vancouver

Biber S, Kannengiesser S, Nistler J, Braun M, Blaess S, Gebhardt M, et al. Design and operation of a whole‐body MRI scanner without RF shielding. Magnetic Resonance in Medicine. 2025;93(4):1842-1855. doi:10.1002/mrm.30374

BibTeX
@article{Biber2025,
  author = {Biber, Stephan and Kannengiesser, Stephan and Nistler, Juergen and Braun, Markus and Blaess, Stefan and Gebhardt, Matthias and Grodzki, David and Ritter, Dieter and Seegerer, Georg and Vester, Markus and Schneider, Rainer},
  title = {Design and operation of a whole‐body MRI scanner without RF shielding},
  journal = {Magnetic Resonance in Medicine},
  year = {2025},
  volume = {93},
  number = {4},
  pages = {1842--1855},
  doi = {10.1002/mrm.30374}
}
Kostenlos starten →