EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Slice‐specific B 1 + shimming improves the repeatability of multishot DWI at 7 T

Belinda Ding iD, Sydney Nicole Williams iD, Iulius Dragonu, Patrick Liebig iD, Sarah Allwood-Spiers, Paul McElhinney, Shajan Gunamony iD, Natasha Fullerton, David Andrew Porter

Magnetic Resonance in Medicine · 2024

DOI: 10.1002/mrm.30208 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2024. Verfasst von Belinda Ding und 8 weiteren Autor:innen. Beteiligte Einrichtung: Siemens Healthcare Ltd, Camberley, UK. Hauptthemen laut MEDLINE: Brain, Diffusion Magnetic Resonance Imaging. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 49 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

Compared with lower field strengths, DWI at 7 T faces the combined challenges of increased distortion and blurring due to B 0 inhomogeneity, and increased signal dropouts due to B 1 + inhomogeneity. This study addresses the B 1 + limitations using slice‐specific static parallel transmission (pTx) in a multi‐shot, readout‐segmented EPI diffusion imaging sequence.

Methods

DWI was performed in 7 healthy subjects using MRI at 7 T and readout‐segmented EPI. Data were acquired with non‐pTx circular‐polarized (CP) pulses (CP‐DWI) and static pTx pulses (pTx‐DWI) using slice‐specific B 1 + shim coefficients. Each volunteer underwent two scan sessions on the same day, with two runs of each sequence in the first session and one run in the second. The sequences were evaluated by assessing image quality, flip‐angle homogeneity, and intrasession and intersession repeatability in ADC estimates. Results pTx‐DWI significantly reduced signal voids compared with CP‐DWI, particularly in inferior brain regions. The use of pTx also improved RF uniformity and symmetry across the brain. These effects translated into improved intrasession and intersession repeatability for pTx‐DWI. Additionally, re‐optimizing the pTx pulse between repeat scans did not have a negative effect on ADC repeatability.

Conclusion

The study demonstrates that pTx provides a reproducible image‐quality increase in multishot DWI at 7 T. The benefits of pTx also extend to quantitative ADC estimation with regard to the improvement in intrasession and intersession repeatability. Overall, the combination of multishot imaging and pTx can support the development of reliable, high‐resolution DWI for clinical studies at 7 T.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

BrainDiffusion Magnetic Resonance ImagingHumansReproducibility of ResultsAdultMaleFemaleEcho-Planar ImagingImage Processing, Computer-AssistedAlgorithmsHealthy VolunteersYoung AdultImage Interpretation, Computer-Assisted

Stichwörter der Autor:innen

RF shimmingdiffusion imagingparallel transmissionrepeatabilityultra‐high‐field

Autor:innen und Institutionen

  • Siemens Healthcare Ltd, Camberley, UK.
  • Imaging Center of Excellence, University of Glasgow, Glasgow, UK.
  • Siemens Healthineers AG, Erlangen, Germany.
  • NHS Greater Glasgow and Clyde, Glasgow, UK.
  • MR CoilTech Ltd, Glasgow, UK.

ORCID-Profile: Belinda Ding, Sydney Nicole Williams, Patrick Liebig, Shajan Gunamony

Zitieren

APA

Ding, B., Williams, S. N., Dragonu, I., Liebig, P., Allwood-Spiers, S., McElhinney, P., Gunamony, S., Fullerton, N., & Porter, D. A. (2024). Slice‐specific B 1 + shimming improves the repeatability of multishot DWI at 7 T. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.30208

Vancouver

Ding B, Williams SN, Dragonu I, Liebig P, Allwood-Spiers S, McElhinney P, et al. Slice‐specific B 1 + shimming improves the repeatability of multishot DWI at 7 T. Magnetic Resonance in Medicine. 2024. doi:10.1002/mrm.30208

BibTeX
@article{Ding2024,
  author = {Ding, Belinda and Williams, Sydney Nicole and Dragonu, Iulius and Liebig, Patrick and Allwood-Spiers, Sarah and McElhinney, Paul and Gunamony, Shajan and Fullerton, Natasha and Porter, David Andrew},
  title = {Slice‐specific B 1 + shimming improves the repeatability of multishot DWI at 7 T},
  journal = {Magnetic Resonance in Medicine},
  year = {2024},
  doi = {10.1002/mrm.30208}
}
Kostenlos starten →