B 0 ‐insensitive image navigators for prospective motion‐corrected MRS with localized second‐order shimming
Magnetic Resonance in Medicine · 2024
DOI: 10.1002/mrm.30151 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2024. Verfasst von Isaac M Adanyeguh und 3 weiteren Autor:innen. Beteiligte Einrichtung: Center for Magnetic Resonance Research and Department of Radiology, University of Minnesota Medical School, Minneapolis, Minnesota, USA. Hauptthemen laut MEDLINE: Motion, Algorithms. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 31 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: NIBIB NIH HHS, NIH HHS, NINDS NIH HHS.
Abstract (Originalsprache)
Purpose
Localized shimming in single‐voxel MRS often results in large B 0 inhomogeneity outside the volume‐of‐interest. This causes unacceptable degradation in motion navigator images. Switching back and forth between whole‐brain shim and localized shim is possible for linear shims, but not for higher‐order shims. Here we propose motion navigators largely insensitive to B 0 inhomogeneity for prospective motion‐corrected MRS with localized higher‐order shimming.
Methods
A recent fast high‐resolution motion navigator based on spiral‐in/out k‐space trajectories and multislice‐to‐volume registration was modified by splitting the readout into multiple shot interleaves which shortened the echo time and reduced the effect of B 0 inhomogeneity. The performance of motion correction was assessed in healthy subjects in the prefrontal cortex using a sLASER sequence at 3T ( N = 5) and 7T ( N = 5).
Results
With multiple spatial interleaves, excellent quality navigator images were acquired in the whole brain in spite of large B 0 inhomogeneity outside the MRS voxel. The total duration of the navigator in sLASER remained relatively short even with multiple shots (3T: 10 spatial interleaves 94 ms per slice; 7T: 15 spatial interleaves 103 ms per slice). Prospective motion correction using the multi‐shot navigators yielded comparable spectral quality (water linewidth and metabolite SNR) with and without subject motion.
Conclusion
B 0 ‐insensitive motion navigators enable prospective motion correction for MRS with all first‐ and second‐order shims adjusted in the MRS voxel, providing optimal spectral linewidth.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Center for Magnetic Resonance Research and Department of Radiology, University of Minnesota Medical School, Minneapolis, Minnesota, USA.
ORCID-Profile: Isaac M Adanyeguh, Dinesh K Deelchand
Förderung
- NIBIB NIH HHS
- NIH HHS
- NINDS NIH HHS
- National Institutes of Health
Zitieren
APA
Adanyeguh, I. M., Park, Y. W., Henry, P. G., & Deelchand, D. K. (2024). B 0 ‐insensitive image navigators for prospective motion‐corrected MRS with localized second‐order shimming. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.30151
Vancouver
Adanyeguh IM, Park YW, Henry PG, Deelchand DK. B 0 ‐insensitive image navigators for prospective motion‐corrected MRS with localized second‐order shimming. Magnetic Resonance in Medicine. 2024. doi:10.1002/mrm.30151
BibTeX
@article{Adanyeguh2024,
author = {Adanyeguh, Isaac M and Park, Young Woo and Henry, Pierre-Gilles and Deelchand, Dinesh K},
title = {B 0 ‐insensitive image navigators for prospective motion‐corrected MRS with localized second‐order shimming},
journal = {Magnetic Resonance in Medicine},
year = {2024},
doi = {10.1002/mrm.30151}
}