Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2024. Verfasst von Dongyue Si und 5 weiteren Autor:innen. Beteiligte Einrichtung: School of Biomedical Engineering and Imaging Sciences, King's College London, London, UK. Hauptthemen laut MEDLINE: Imaging, Three-Dimensional, Phantoms, Imaging, Adipose Tissue. Die Arbeit wurde bisher 11-mal zitiert (Crossref) und nennt 50 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Wellcome EPSRC Centre for Medical Engineering, the Department of Health through the National Institute for Health Research (NIHR) comprehensive Biomedical Research Centre award, ANID Millenium Institute for Intelligent Healthcare Engineering.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

To develop and validate a highly efficient motion compensated free‐breathing isotropic resolution 3D whole‐heart joint T 1 /T 2 mapping sequence with anatomical water/fat imaging at 0.55 T.

Methods

The proposed sequence takes advantage of shorter T 1 at 0.55 T to acquire three interleaved water/fat volumes with inversion‐recovery preparation, no preparation, and T 2 preparation, respectively. Image navigators were used to facilitate nonrigid motion‐compensated image reconstruction. T 1 and T 2 maps were jointly calculated by a dictionary matching method. Validations were performed with simulation, phantom, and in vivo experiments on 10 healthy volunteers and 1 patient. The performance of the proposed sequence was compared with conventional 2D mapping sequences including modified Look‐Locker inversion recovery and T 2 ‐prepared balanced steady‐SSFP sequence.

Results

The proposed sequence has a good T 1 and T 2 encoding sensitivity in simulation, and excellent agreement with spin‐echo reference T 1 and T 2 values was observed in a standardized T 1 /T 2 phantom ( R 2 = 0.99). In vivo experiments provided good‐quality co‐registered 3D whole‐heart T 1 and T 2 maps with 2‐mm isotropic resolution in a short scan time of about 7 min. For healthy volunteers, left‐ventricle T 1 mean and SD measured by the proposed sequence were both comparable with those of modified Look‐Locker inversion recovery (640 ± 35 vs. 630 ± 25 ms [ p = 0.44] and 49.9 ± 9.3 vs. 54.4 ± 20.5 ms [ p = 0.42]), whereas left‐ventricle T 2 mean and SD measured by the proposed sequence were both slightly lower than those of T 2 ‐prepared balanced SSFP (53.8 ± 5.5 vs. 58.6 ± 3.3 ms [ p < 0.01] and 5.2 ± 0.9 vs. 6.1 ± 0.8 ms [ p = 0.03]). Myocardial T 1 and T 2 in the patient measured by the proposed sequence were in good agreement with conventional 2D sequences and late gadolinium enhancement.

Conclusion

The proposed sequence simultaneously acquires 3D whole‐heart T 1 and T 2 mapping with anatomical water/fat imaging at 0.55 T in a fast and efficient 7‐min scan. Further investigation in patients with cardiovascular disease is now warranted.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Imaging, Three-DimensionalPhantoms, ImagingAdipose TissueHumansMaleAdultHeartReproducibility of ResultsAlgorithmsFemaleMagnetic Resonance ImagingRespirationBody WaterImage Interpretation, Computer-AssistedImage Processing, Computer-AssistedHealthy Volunteers

Stichwörter der Autor:innen

cardiac MRIfree breathinglow fieldmyocardial T1 mappingmyocardial T2 mappingthree dimensional

Autor:innen und Institutionen

  • School of Biomedical Engineering and Imaging Sciences, King's College London, London, UK.
  • MR Research Collaborations, Siemens Healthcare Limited, Camberley, UK.
  • School of Engineering, Pontificia Universidad Católica de Chile, Santiago, Chile.
  • Millennium Institute for Intelligent Healthcare Engineering, Santiago, Chile.
  • Institute of Biological and Medical Engineering, Pontificia Universidad Católica de Chile, Santiago, Chile.
  • British Heart Foundation Centre of Research Excellence, King's College London, London, UK.

ORCID-Profile: Michael G Crabb, Claudia Prieto, René M Botnar

Förderung

  • Wellcome EPSRC Centre for Medical Engineering
  • the Department of Health through the National Institute for Health Research (NIHR) comprehensive Biomedical Research Centre award
  • ANID Millenium Institute for Intelligent Healthcare Engineering
  • EPSRC
  • ANID IMPACT, Basal Center of Interventional Medicine for Precision and Advanced Cellular Therapy
  • the Technical University of Munich - Institute for Advanced Study
  • National Institute for Health and Research Cardiovascular MedTech Co-operative
  • British Heart Foundation

Zitieren

APA

Si, D., Crabb, M. G., Kunze, K. P., Littlewood, S. J., Prieto, C., & Botnar, R. M. (2024). Free‐breathing 3D whole‐heart joint T 1 / T 2 mapping and water/fat imaging at 0. 55 T. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.30139

Vancouver

Si D, Crabb MG, Kunze KP, Littlewood SJ, Prieto C, Botnar RM. Free‐breathing 3D whole‐heart joint T 1 / T 2 mapping and water/fat imaging at 0. 55 T. Magnetic Resonance in Medicine. 2024. doi:10.1002/mrm.30139

BibTeX
@article{Si2024,
  author = {Si, Dongyue and Crabb, Michael G and Kunze, Karl P and Littlewood, Simon J and Prieto, Claudia and Botnar, René M},
  title = {Free‐breathing 3D whole‐heart joint T 1 / T 2 mapping and water/fat imaging at 0. 55 T},
  journal = {Magnetic Resonance in Medicine},
  year = {2024},
  doi = {10.1002/mrm.30139}
}
Kostenlos starten →