Quantification of water exchange across the blood‐brain barrier using noncontrast MR fingerprinting
Magnetic Resonance in Medicine · 2024
DOI: 10.1002/mrm.30127 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2024. Verfasst von Emma L Thomson und 3 weiteren Autor:innen. Beteiligte Einrichtung: Centre for Medical Image Computing, Department of Medical Physics and Biomedical Engineering, University College London, London, UK. Hauptthemen laut MEDLINE: Magnetic Resonance Imaging, Blood-Brain Barrier, Computer Simulation, Water. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 55 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: The National Institute for Health and Care Research University College London Hospitals Biomedical Research Centre, EU Horizon2020 Research and Innovation Action Grants Human Brain Project, Rosetrees Trust.
Abstract (Originalsprache)
Purpose
A method is proposed to quantify cerebral blood volume () and intravascular water residence time () using MR fingerprinting (MRF), applied using a spoiled gradient echo sequence without the need for contrast agent.
Methods
An in silico study optimized an acquisition protocol to maximize the sensitivity of the measurement to and changes. Its accuracy in the presence of variations in , , and was evaluated. The optimized protocol (scan time of 19 min) was then tested in a exploratory healthy volunteer study (10 volunteers, mean age 24 3, six males) at 3 T with a repeat scan taken after repositioning to allow estimation of repeatability.
Results
Simulations show that assuming literature values for and , no variation in , while fitting only and , leads to large errors in quantification of and , regardless of noise levels. However, simulations also show that matching , , , and , simultaneously is feasible at clinically achievable noise levels. Across the healthy volunteers, all parameter quantifications fell within the expected literature range. In addition, the maps show good agreement between hemispheres suggesting physiologically relevant information is being extracted. Expected differences between white and gray matter ( p < 0.0001) and ( p < 0.0001) are observed, and show no significant differences, p = 0.4 and p = 0.6, respectively. Moderate to excellent repeatability was seen between repeat scans: mean intra‐class correlation coefficient of , , , and .
Conclusion
We demonstrate that regional simultaneous quantification of , , , and using MRF is feasible in vivo.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Centre for Medical Image Computing, Department of Medical Physics and Biomedical Engineering, University College London, London, UK.
- NMR Research Unit, Queen Square MS Centre, Department of Neuroinflammation, UCL Queen Square Institute of Neurology, Faculty of Brain Sciences, London, UK.
- Department of Brain & Behavioural Sciences, University of Pavia, Pavia, Italy.
- IRCCS Mondino Foundation, Pavia, Italy.
- Bioxydyn Limited, Manchester, UK.
ORCID-Profile: Emma L Thomson, Elizabeth Powell, Claudia A M Gandini Wheeler-Kingshott, Geoff J M Parker
Förderung
- The National Institute for Health and Care Research University College London Hospitals Biomedical Research Centre
- EU Horizon2020 Research and Innovation Action Grants Human Brain Project
- Rosetrees Trust
- EPSRC
- EPSRC Centre for Doctoral Training in Intelligent, Integrated Imaging In Healthcare (i4health)
- MRC
- Engineering and Physical Sciences Research Council
Zitieren
APA
Thomson, E. L., Powell, E., Gandini Wheeler-Kingshott, C. A. M., & Parker, G. J. M. (2024). Quantification of water exchange across the blood‐brain barrier using noncontrast MR fingerprinting. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.30127
Vancouver
Thomson EL, Powell E, Gandini Wheeler-Kingshott CAM, Parker GJM. Quantification of water exchange across the blood‐brain barrier using noncontrast MR fingerprinting. Magnetic Resonance in Medicine. 2024. doi:10.1002/mrm.30127
BibTeX
@article{Thomson2024,
author = {Thomson, Emma L and Powell, Elizabeth and Gandini Wheeler-Kingshott, Claudia A M and Parker, Geoff J M},
title = {Quantification of water exchange across the blood‐brain barrier using noncontrast MR fingerprinting},
journal = {Magnetic Resonance in Medicine},
year = {2024},
doi = {10.1002/mrm.30127}
}