EUVIMEDEUVIMED entdecken
Open Access · CC BY-NC 4.0

Multiple‐ TE Based Blood–Brain‐Barrier Water Exchange Time Measurement Using a TE ‐Resolved 3D TSE Stack‐Of‐Spirals Readout

Bo Li iD, Yiran Li, Xiao Liang iD, Amnah Mahroo iD, Manuel Taso, Marta Vidorreta Díaz de Cerio, Lucas Lemos Franco, Yulin Chang, María Asunción Fernández Seara, Matthias Günther iD, John A Detre, Ze Wang

Magnetic Resonance in Medicine · 2026

DOI: 10.1002/mrm.70450 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2026. Verfasst von Bo Li und 11 weiteren Autor:innen. Beteiligte Einrichtung: Department of Diagnostic Radiology and Nuclear Medicine, University of Maryland, Baltimore, Maryland, USA. Hauptthemen laut MEDLINE: Blood-Brain Barrier, Imaging, Three-Dimensional, Magnetic Resonance Imaging. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 75 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0). Gefördert von: NIA NIH HHS, NIBIB NIH HHS, National Institute of Biomedical Imaging and Bioengineering.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

To develop a slice‐wise blurring‐free and densely sampled TE‐resolved multiple‐TE (mTE) ASL sequence (TASL) for measuring blood–brain barrier (BBB) water exchange time.

Methods

A 3D TSE spiral‐readout pCASL sequence was modified to enable TE‐resolved acquisition. Signals at mTEs from a single kz partition were acquired within a single shot, and the acquisition was repeated until all required kz partitions were covered. In vivo experiments were conducted on healthy volunteers using both a conventional pCASL sequence and the proposed TASL sequence. Signal intensity modulation, perfusion SNR, and quantitative parameters including cerebral blood flow (CBF), arterial transit time (ATT), intravoxel transit time ( ITT ), and water exchange time ( T exch ) were evaluated. Test–retest scans were conducted to assess reproducibility.

Results

Compared with the conventional mTE acquisition, TASL eliminates T 2 decay‐induced slice‐wise intensity modulation and blurring, yielding higher whole‐brain perfusion SNR across all TEs and PLDs. Estimates of CBF and ATT showed negligible differences between the two methods (both < 1%), whereas mean T exch and ITT values were 18.2% and 9.05% higher, respectively, with TASL. Bland–Altman analysis showed minimal bias, with 95% of differences within the limits of agreement (mean±1.96 SD). The inter‐session ICC and wsCV of CBF, ATT, T exch , and ITT in gray matter are 0.86% and 8.38%, 0.96% and 1.98%, 0.93% and 4.04%, and 0.95% and 4.32%, respectively.

Conclusion

TASL eliminates signal intensity modulation and blurring artifacts while providing densely sampled TE for parametric fitting. It may be a valuable tool for accurate and reliable quantification of cerebral perfusion and BBB permeability.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Blood-Brain BarrierImaging, Three-DimensionalMagnetic Resonance ImagingHumansReproducibility of ResultsMaleAdultCerebrovascular CirculationBrainWaterFemaleAlgorithmsSignal-To-Noise RatioHealthy Volunteers

Stichwörter der Autor:innen

BBBTE‐resolved readoutmultiple‐TEpCASLspiral

Substanzen

Water

Autor:innen und Institutionen

  • Department of Diagnostic Radiology and Nuclear Medicine, University of Maryland, Baltimore, Maryland, USA.
  • Fraunhofer Institute for Digital Medicine MEVIS, Bremen, Germany.
  • Krembil Brain Institute, University Health Network, Toronto, Canada.
  • Department of Radiology, Perelman School of Medicine, University of Pennsylvania, Pennsylvania, USA.
  • Siemens Healthcare, Madrid, Spain.
  • Siemens Medical Solutions USA Inc, Malvern, Pennsylvania, USA.

ORCID-Profile: Bo Li, Xiao Liang, Amnah Mahroo, Matthias Günther

Förderung

  • NIA NIH HHS
  • NIBIB NIH HHS
  • National Institute of Biomedical Imaging and Bioengineering
  • National Institute on Aging

Zitieren

APA

Li, B., Li, Y., Liang, X., Mahroo, A., Taso, M., de Cerio, M. V. D., Franco, L. L., Chang, Y., Seara, M. A. F., Günther, M., Detre, J. A., & Wang, Z. (2026). Multiple‐ TE Based Blood–Brain‐Barrier Water Exchange Time Measurement Using a TE ‐Resolved 3D TSE Stack‐Of‐Spirals Readout. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.70450

Vancouver

Li B, Li Y, Liang X, Mahroo A, Taso M, de Cerio MVD, et al. Multiple‐ TE Based Blood–Brain‐Barrier Water Exchange Time Measurement Using a TE ‐Resolved 3D TSE Stack‐Of‐Spirals Readout. Magnetic Resonance in Medicine. 2026. doi:10.1002/mrm.70450

BibTeX
@article{Li2026,
  author = {Li, Bo and Li, Yiran and Liang, Xiao and Mahroo, Amnah and Taso, Manuel and de Cerio, Marta Vidorreta Díaz and Franco, Lucas Lemos and Chang, Yulin and Seara, María Asunción Fernández and Günther, Matthias and Detre, John A and Wang, Ze},
  title = {Multiple‐ TE Based Blood–Brain‐Barrier Water Exchange Time Measurement Using a TE ‐Resolved 3D TSE Stack‐Of‐Spirals Readout},
  journal = {Magnetic Resonance in Medicine},
  year = {2026},
  doi = {10.1002/mrm.70450}
}
Kostenlos starten →