EUVIMEDEUVIMED entdecken
Open Access

Feasibility of free‐breathing 19 F MRI image acquisition to characterize ventilation defects in CF and healthy volunteers at wash‐in

Sang Hun Chung iD, Khoi Minh Huynh, Jennifer L Goralski, Yong Chen, Pew-Thian Yap, Agathe S Ceppe, Margret Z Powell, Scott H Donaldson, Yueh Z Lee iD

Magnetic Resonance in Medicine · 2023

DOI: 10.1002/mrm.29630 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2023. Verfasst von Sang Hun Chung und 8 weiteren Autor:innen. Beteiligte Einrichtung: Department of Biomedical Engineering, University of North Carolina at Chapel Hill, Chapel Hill, North Carolina, USA. Hauptthemen laut MEDLINE: Cystic Fibrosis, Respiration Disorders. Die Arbeit wurde bisher 8-mal zitiert (Crossref) und nennt 59 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: NHLBI NIH HHS, NIDDK NIH HHS, Cystic Fibrosis Foundation.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

To explore the feasibility of measuring ventilation defect percentage (VDP) using 19 F MRI during free‐breathing wash‐in of fluorinated gas mixture with postacquisition denoising and to compare these results with those obtained through traditional Cartesian breath‐hold acquisitions.

Methods

Eight adults with cystic fibrosis and 5 healthy volunteers completed a single MR session on a Siemens 3T Prisma. 1 H Ultrashort‐TE MRI sequences were used for registration and masking, and ventilation images with 19 F MRI were obtained while the subjects breathed a normoxic mixture of 79% perfluoropropane and 21% oxygen (O 2 ). 19 F MRI was performed during breath holds and while free breathing with one overlapping spiral scan at breath hold for VDP value comparison. The 19 F spiral data were denoised using a low‐rank matrix recovery approach.

Results

VDP measured using 19 F VIBE and 19 F spiral images were highly correlated ( r = 0.84) at 10 wash‐in breaths. Second‐breath VDPs were also highly correlated ( r = 0.88). Denoising greatly increased SNR (pre‐denoising spiral SNR, 2.46 ± 0.21; post‐denoising spiral SNR, 33.91 ± 6.12; and breath‐hold SNR, 17.52 ± 2.08).

Conclusion

Free‐breathing 19 F lung MRI VDP analysis was feasible and highly correlated with breath‐hold measurements. Free‐breathing methods are expected to increase patient comfort and extend ventilation MRI use to patients who are unable to perform breath holds, including younger subjects and those with more severe lung disease.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Cystic FibrosisRespiration DisordersAdultHumansHealthy VolunteersFeasibility StudiesRespirationLungMagnetic Resonance ImagingOxygen

Stichwörter der Autor:innen

19FVDPcystic fibrosisfree breathinglung MRI

Substanzen

Oxygen

Autor:innen und Institutionen

  • Department of Biomedical Engineering, University of North Carolina at Chapel Hill, Chapel Hill, North Carolina, USA.
  • Division of Pulmonary and Critical Care Medicine, University of North Carolina at Chapel Hill, Chapel Hill, North Carolina, USA.
  • Marsico Lung Institute/UNC Cystic Fibrosis Center, University of North Carolina at Chapel Hill, Chapel Hill, North Carolina, USA.
  • Division of Pediatric Pulmonology, University of North Carolina at Chapel Hill, Chapel Hill, North Carolina, USA.
  • Department of Radiology, Case Western Reserve University, Cleveland, Ohio, USA.
  • Department of Radiology and Biomedical Research Imaging Center, University of North Carolina at Chapel Hill, Chapel Hill, North Carolina, USA.

ORCID-Profile: Sang Hun Chung, Yueh Z Lee

Förderung

  • NHLBI NIH HHS
  • NIDDK NIH HHS
  • Cystic Fibrosis Foundation
  • Korea National Institute of Health

Zitieren

APA

Chung, S. H., Huynh, K. M., Goralski, J. L., Chen, Y., Yap, P. T., Ceppe, A. S., Powell, M. Z., Donaldson, S. H., & Lee, Y. Z. (2023). Feasibility of free‐breathing 19 F MRI image acquisition to characterize ventilation defects in CF and healthy volunteers at wash‐in. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.29630

Vancouver

Chung SH, Huynh KM, Goralski JL, Chen Y, Yap PT, Ceppe AS, et al. Feasibility of free‐breathing 19 F MRI image acquisition to characterize ventilation defects in CF and healthy volunteers at wash‐in. Magnetic Resonance in Medicine. 2023. doi:10.1002/mrm.29630

BibTeX
@article{Chung2023,
  author = {Chung, Sang Hun and Huynh, Khoi Minh and Goralski, Jennifer L and Chen, Yong and Yap, Pew-Thian and Ceppe, Agathe S and Powell, Margret Z and Donaldson, Scott H and Lee, Yueh Z},
  title = {Feasibility of free‐breathing 19 F MRI image acquisition to characterize ventilation defects in CF and healthy volunteers at wash‐in},
  journal = {Magnetic Resonance in Medicine},
  year = {2023},
  doi = {10.1002/mrm.29630}
}
Kostenlos starten →