Evaluation Of A Semi-Automated Indirect Immunofluorescence Assay Using Validated In-House Hep-2 Cell Slides As A Reference Standard For Antinuclear Antibody Detection

Muhammad Zain Arshad, Muhammad Omair Riaz, Muhammad Hussain, Dawood Ahmad, Ayesha Tanveer, Muhammad Aftab Hassan

Pakistan Armed Forces Medical Journal · 2026-10-30 · 76 (6) · 775-779

DOI: 10.51253/pafmj.v76i6.10690 ↗
AbstractBibliografische Angaben

Abstract

Objective: To evaluate the analytical performance of a semi-automated assay for detection of antinuclear antibodies by indirect immunofluorescence. Study Design: Cross-sectional study Place and Duration of Study: Department of Immunology, Armed Forces Institute of Pathology (AFIP), Rawalpindi, Pakistan, from Jan to Jun 2023 Methodology: A total of thirty-six serum samples were selected from patients with antinuclear antibody-positive systemic autoimmune rheumatic diseases (n = 26) and from healthy controls (n = 10). Samples were applied on in-house HEp-2 cell slides and then subsequently stained and interpreted by two experienced observers independently. In parallel, ANA tests were also performed using a semi-automated indirect immunofluorescence assay (IFA) benchtop analyzer (Akiron NEO) and software using a standard kit as per the manufacturer's instructions. Positive/negative results, staining patterns, and intensities were compared between in-house HEp-2 cell slides as a reference standard and the semi-automated assay. Relative sensitivities, specificities, and degree of concordance were assessed by overall percentage agreement or by Cohen’s k. Results: Among a total of 36 samples, in-house HEp-2 cell slides yielded 26 (72.22%) positive and 10 (27.78%) negative. The semi-automated IFA system yielded 21(58.33%) positive samples and 15(41.66%) negative samples. Overall relative sensitivity, specificity, and concordance of the semi-automated IFA system were 65.38%, 60%, and 63.88% (k = 0.21), respectively (p = 0.26). Conclusions: The semi-automated IFA system (Akiron NEO) has fair agreement with in-house HEp-2 cell slides with respect to positivity/negativity and pattern recognition but poor agreement for intensity. This makes semi-automated evaluation of ANA relatively less suitable for routine testing.

Wissen, das zu Ihrer Arbeit passt.

Vom kostenlosen Einstieg über die persönliche Recherche bis zur vollständigen Research Suite.

Tarife ansehen →

Sind alle Verlagsvolltexte enthalten?

Nein. Bibliografische Nachweise, Abstracts und Volltexte sind unterschiedliche Inhalte. Ob ein Volltext verfügbar ist, hängt von der Quelle, Open Access und Ihren vorhandenen Lizenzen ab. Ein EUVIMED-Tarif ersetzt keine zusätzlichen Verlagsrechte.

Tarife ansehen →