An End-to-end Quality Assurance Procedure for Ethos Online Adaptive Radiotherapy

S. A. Yoganathan, Amine Khemissi, Satheesh Paloor, Rabih Hammoud, Noora Al-Hammadi

Journal of Medical Physics · 2025-01 · 50 (1) · 140-147

DOI: 10.4103/jmp.jmp_188_24 ↗
AbstractBibliografische Angaben

Abstract

Abstract Purpose: Online adaptive radiation therapy (OART) poses unique challenges for quality assurance (QA), requiring innovative methodologies beyond traditional techniques. This study introduced an end-to-end (E2E) QA test for the Ethos OART system. Materials and Methods: Initial treatment plans were developed using deformed computed tomography (CT) images of standard phantoms. During treatment sessions, adaptive plans were created and delivered using undistorted physical QA phantoms equipped with measuring detectors. Our approach was demonstrated using standard QA phantoms – OCTAVIUS-four-dimensional (PTW, Freiburg, Germany), ArcCHECK (Sun Nuclear Corp., FL, USA), and the RUBY (PTW, Freiburg, Germany) – to evaluate the accuracy of contouring, synthetic CT (sCT), and dosimetry of adaptive plans in the Ethos OART system. Results: Our findings demonstrated the superior performance of the Ethos OART system, with a gamma pass rate exceeding 96% (2% local/2 mm) and point dose deviations below 0.5%. The Dice coefficients for body contours between the sCT and reference CT were above 0.9, and the sCT accuracy was confirmed by mean absolute errors of

Wissen, das zu Ihrer Arbeit passt.

Vom kostenlosen Einstieg über die persönliche Recherche bis zur vollständigen Research Suite.

Tarife ansehen →

Sind alle Verlagsvolltexte enthalten?

Nein. Bibliografische Nachweise, Abstracts und Volltexte sind unterschiedliche Inhalte. Ob ein Volltext verfügbar ist, hängt von der Quelle, Open Access und Ihren vorhandenen Lizenzen ab. Ein EUVIMED-Tarif ersetzt keine zusätzlichen Verlagsrechte.

Tarife ansehen →