Magnetic resonance imaging-induced forces on video capsule endoscopes: First systematic in vitro assessment
World Journal of Gastroenterology · 2027-01-07 · 33 (1)
DOI: 10.3748/wjg.122550 ↗Abstract
BACKGROUND Magnetic resonance imaging (MRI) is contraindicated in patients with an ingested video capsule endoscopy (VCE) due to the theoretical risk of mechanical and thermal harm. However, isolated cases of inadvertent scanning have not shown clear harm. Manufacturers advise against performing VCE in patients with an MRI scheduled within the next 14 days, yet asymptomatic retention for up to 12 years has been reported. With the increasing use of VCEs, the likelihood of incidental MRI exposure is rising. To date, no systematic data have been published on the forces exerted on VCEs in an MRI environment. AIM To quantify magnetically induced forces on VCEs in a 3-Tesla (3-T) scanner to assess clinical safety risks. METHODS Displacement forces on 11 commercially available VCE models were measured in a 3-T MRI scanner according to the American Society of Testing Materials (ASTM) F2052-21. Following ASTM F2213-17, torque was measured by the pulley method for eight VCEs and calculated for three magnet-containing gastric capsules with high displacement forces. The lengths of internal metallic components of the VCEs, as indicators of potential thermal effects, were determined using X-ray images. RESULTS VCEs containing magnets showed significantly higher median displacement forces than those without magnets (2366 mN vs 293.92 mN, P = 0.012). Magnetically induced displacement forces were 50.7-fold higher than the individual gravitational force for magnet-containing devices (2366 mN vs 46.66 mN), compared to a 9.7-fold excess for capsules without a magnet (293.92 mN vs 30.20 mN). Median torque was 19.2 mNm for capsules without integrated magnets vs 289.02 mNm with magnets (P = 0.017). For all but one VCE, the forces exceeded the safety threshold defined by the capsule’s gravitational force. The mean length of internal metallic components was 21.3 mm (range: 18.2-25.4 mm); only two capsules slightly exceeded the 24-mm threshold for possible thermal effects. CONCLUSION In a 3-T MRI, all but one VCE, particularly magnet-containing capsules, exceeded accepted safety thresholds for displacement force and torque but remained below perforation-relevant levels.
EUVIMED