Color Stability of Vita 3D-Master Shade Tabs Following Repeated Disinfection: An In Vitro Study
Jordan Journal of Dentistry · 2027
DOI: 10.14525/jjd.90 ↗Abstract
Objectives: To evaluate the color stability of VITA 3D-Master shade tabs following repeated disinfection with Peroxan 6% and Desomed 0.5%, compared to a distilled water control. Materials and Methods: Ninety VITA 3D-Master shade tabs representing 10 shades were divided into 3 groups (n = 30): Control, Peroxan 6%, and Desomed 0.5%. Each tab was immersed for 10 minutes per cycle and subjected to 200 disinfection cycles. Color measurements were performed at baseline, 50, 100, 150, and 200 cycles using a VITA Easyshade V spectrophotometer. Color differences were calculated using the CIEDE2000 formula. Data was analyzed using repeated-measures ANOVA and Tukey’s test (α = 0.05). Results: At 200 cycles, mean ΔE00 values were 0.9 ± 0.2 for Control, 2.8 ± 0.4 for Peroxan, and 3.6 ± 0.5 for Desomed tabs. 20% of Peroxan tabs and 100% of Desomed tabs exceeded the clinical acceptability threshold of ΔE00 > 2.7. Repeated-measures two-way ANOVA showed significant main effects for group (p < 0.001), time (p < 0.001) and group × time interaction (p
EUVIMED