Acetabular Debonding: An Investigation of Porous Coating Delamination in Hip Resurfacing Arthroplasty
Advances in Orthopedics · 2018-11-01 · 2018 · 1-10
DOI: 10.1155/2018/5282167 ↗Abstract
Background. To date, there have been no published investigations on the cause of acetabular debonding, a rare failure phenomenon in metal-on-metal hip resurfacing where the acetabular porous coating delaminates from the implant while remaining well fixed to the pelvic bone. Purposes. This study aims to summarize the current understanding of acetabular debonding and to investigate the discrepancy in rate of debonding between two implant systems. Patients and Methods. To elucidate potential causes of debonding, we retrospectively analyzed a single-surgeon cohort of 839 hip resurfacing cases. Specifically, we compared rate of debonding and manufacturing processes between two implant systems. Results. Group 1 experienced significantly more cases of debonding than Group 2 cases (4.0% versus 0.0%, p value