Pulse Oximetry for the Detection of Obstructive Sleep Apnea Syndrome: Can the Memory Capacity of Oxygen Saturation Influence Their Diagnostic Accuracy?
Sleep Disorders · 2011
DOI: 10.1155/2011/427028 ↗Abstract
Objective . To assess the diagnostic ability of WristOx 3100 using its three different recording settings in patients with suspected obstructive sleep apnea syndrome (OSAS). Methods . All participants (135) performed the oximetry (three oximeters WristOx ) and polysomnography (PSG) simultaneously in the sleep laboratory. Both recordings were interpreted blindly. Each oximeter was set to one of three different recording settings (memory capabilities 0.25, 0.5, and 1 Hz). The software (nVision 5.1) calculated the adjusted O 2 desaturation index-mean number of O 2 desaturation per hour of analyzed recording ≥2, 3, and 4% (ADI2, 3, and 4). The ADI2, 3, and 4 cutoff points that better discriminated between subjects with or without OSAS arose from the receiver-operator characteristics (ROCs) curve analysis. OSAS was defined as a respiratory disturbance index (RDI) ≥ 5. Results . 101 patients were included (77 men, mean age 52, median RDI 22.6, median BMI 27.4 kg/m 2 ). The area under the ROCs curves (AUC-ROCs) of ADI2, 3, and 4 with different data storage rates were similar (AUC-ROCs with data storage rates of 0.25/0.5/1 Hz: ADI2: 0.958/0.948/0.965, ADI3: 0.961/0.95/0.966, and ADI4: 0.957/0.949/0.963, P NS). Conclusions . The ability of WristOx 3100 to detect patients with OSAS was not affected by the data storage rate of the oxygen saturation signal. Both memory capacity of 0.25, 0.5, or 1 Hz showed a similar performance for the diagnosis of OSAS.