Analytical Validation of β‐Hydroxybutyrate Quantification in Feline Serum with the BT3500 Analyzer

Alessandro Tirolo iD, Maria Chiara Sabetti, Alicia Maria Carrillo Heredero iD, Francesca Fidanzio, Andrea Corsini iD

Veterinary Clinical Pathology · 11.11.2026

DOI: 10.1111/vcp.70163 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Veterinary Clinical Pathology, online erschienen am 11.11.2026. Verfasst von Alessandro Tirolo und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Veterinary Science, University of Parma, Parma, Italy. Die Arbeit nennt 16 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Quantitative measurement of serum β‐hydroxybutyrate (β‐OHB) is essential for detecting and monitoring ketosis and ketoacidosis in diabetic cats, particularly in the era of sodium–glucose cotransporter two (SGLT2) inhibitors, where euglycemic ketoacidosis may occur. The BT3500 analyzer has not been analytically validated for β‐OHB quantification in feline serum.

Objectives

Analytically validate the Randox Ranbut enzymatic β‐OHB assay on feline serum using the BT3500 chemistry analyzer, including evaluation of sample stability under refrigerated and frozen storage conditions.

Methods

Residual feline serum samples and pooled sera were used to assess intra‐assay and inter‐assay precision, spike–recovery accuracy across five concentration increments, linearity under serial dilution, and analytical sensitivity (LoB, LoD, LoQ). Short‐term stability was evaluated at 4°C for 7 days and long‐term stability at −20°C at predefined time points up to 360 days.

Results

Intra‐assay coefficient of variation (CV%) ranged from 1.68% to 14.48% and inter‐assay CV% from 3.08% to 10.20%, with higher imprecision at low concentrations. Recoveries ranged from 110% to 122% (mean 114% ± 4%). Linearity was excellent (slope 1.012; R 2 ≥ 0.9998). LoB, LoD, and LoQ were 0.044, 0.091, and 0.150 mmol/L, respectively. β‐OHB was stable at 4°C up to 6 days and at −20°C up to 180 days, while decreases exceeding the acceptance limit were observed at 360 days.

Conclusions

The Randox β‐OHB assay on the BT3500 demonstrated acceptable analytical performance in feline serum and is suitable for routine clinical use; frozen storage beyond 180 days may impact result accuracy.

Schlagwörter

Stichwörter der Autor:innen

catsdiabetes mellitusketone bodiesketosismethod validation

Autor:innen und Institutionen

  • Department of Veterinary Science, University of Parma, Parma, Italy.

ORCID-Profile: Alessandro Tirolo, Alicia Maria Carrillo Heredero, Andrea Corsini

Zitieren

APA

Tirolo, A., Sabetti, M. C., Carrillo Heredero, A. M., Fidanzio, F., & Corsini, A. (2026). Analytical Validation of β‐Hydroxybutyrate Quantification in Feline Serum with the BT3500 Analyzer. Veterinary Clinical Pathology. https://doi.org/10.1111/vcp.70163

Vancouver

Tirolo A, Sabetti MC, Carrillo Heredero AM, Fidanzio F, Corsini A. Analytical Validation of β‐Hydroxybutyrate Quantification in Feline Serum with the BT3500 Analyzer. Veterinary Clinical Pathology. 2026. doi:10.1111/vcp.70163

BibTeX
@article{Tirolo2026,
  author = {Tirolo, Alessandro and Sabetti, Maria Chiara and Carrillo Heredero, Alicia Maria and Fidanzio, Francesca and Corsini, Andrea},
  title = {Analytical Validation of β‐Hydroxybutyrate Quantification in Feline Serum with the BT3500 Analyzer},
  journal = {Veterinary Clinical Pathology},
  year = {2026},
  doi = {10.1111/vcp.70163}
}
Kostenlos starten →