Assessment of accuracy of two pulse oximeters in infants with cyanotic and acyanotic congenital heart diseases
Pediatric Anesthesia · 2024
DOI: 10.1111/pan.15017 ↗Kurz gefasst
Fachartikel in der Zeitschrift Pediatric Anesthesia, 2024. Verfasst von Yordan Hristov Georgiev und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Pediatric Cardiology, Pulmonology and Pediatric Intensive Care Medicine, University Children's Hospital Tübingen, Tübingen, Germany. Hauptthemen laut MEDLINE: Oximetry, Heart Defects, Congenital, Cyanosis, Oxygen Saturation. Die Arbeit nennt 19 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background
Peripherally measured oxygen saturation ( SpO 2 ) may often differ from arterial oxygen saturation ( SaO 2 ), measured by co‐oximetry, especially within the lower range of oxygen saturations. This can potentially impact clinical decisions and therapy in children with congenital heart disease, as critical hypoxemia might remain unnoticed.
Aims
Our aim was to investigate the accuracy of two different pulse oximeters compared to SaO 2 in infants with congenital heart diseases.
Methods
Simultaneous recordings of SpO 2 , measured by two different pulse oximeters (Philips IntelliVue X3 Monitor and Nellcor™ OxiMax™), were compared to SaO 2 obtained by arterial blood gas analysis.
Results
A total of 153 measurements were performed in 44 infants with arterial oxygen saturation between 70 and 100%. We divided the measurements into 3 subgroups: group 1—SaO 2 70.0%–85.0%, group 2—SaO 2 85.1%–94.0%, group 3—SaO 2 >94.1%. For Philipps, the median bias was 5.3 (IQR: 2.6–8.7) %, 2.3 (IQR: 0.9–6.0) % and 1.1 (IQR: −0.8–2.4) % in group 1, 2 and 3, respectively. For OxiMax™, the median bias was 2.7 (IQR: 0.5–5.1) %, 0.2 (IQR: −0.9–2.6) % and −0.5 (IQR: −1.3–0.6) % in group 1, 2 and 3, respectively. Regarding the accuracy of these oximeters, as evaluated with the Accuracy root mean squared index (Arms), it was 9.8 versus 4.5% in group 1, 4.5 versus 2.9% in group 2 and 2.4 versus 1.9% in group 3 for Philipps and OxiMax™, respectively.
Conclusions
In lower range saturations between 70% and 85% the accuracy of both pulse oximeters exceeded the threshold of ≤3% recommended by the Food and Drug Administration (FDA). Therefore, peripheral pulse oximetry within the lower range of oxygen saturations should be interpreted with caution in infants with congenital heart diseases, taking into consideration its limitations. Direct co‐oximetry should be the preferred method to support clinical decisions in children with cyanotic congenital heart diseases.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Pediatric Cardiology, Pulmonology and Pediatric Intensive Care Medicine, University Children's Hospital Tübingen, Tübingen, Germany.
ORCID-Profile: Yordan Hristov Georgiev, Jörg Michel
Zitieren
APA
Georgiev, Y. H., Neunhoeffer, F., Hofbeck, M., & Michel, J. (2024). Assessment of accuracy of two pulse oximeters in infants with cyanotic and acyanotic congenital heart diseases. Pediatric Anesthesia. https://doi.org/10.1111/pan.15017
Vancouver
Georgiev YH, Neunhoeffer F, Hofbeck M, Michel J. Assessment of accuracy of two pulse oximeters in infants with cyanotic and acyanotic congenital heart diseases. Pediatric Anesthesia. 2024. doi:10.1111/pan.15017
BibTeX
@article{Georgiev2024,
author = {Georgiev, Yordan Hristov and Neunhoeffer, Felix and Hofbeck, Michael and Michel, Jörg},
title = {Assessment of accuracy of two pulse oximeters in infants with cyanotic and acyanotic congenital heart diseases},
journal = {Pediatric Anesthesia},
year = {2024},
doi = {10.1111/pan.15017}
}
EUVIMED