Pacing and Ablation Technique Using Microelectrode for Pulmonary Vein Isolation Using a Local Impedance‐Guided Catheter
Pacing and Clinical Electrophysiology · 2025
DOI: 10.1111/pace.15144 ↗Kurz gefasst
Übersichtsarbeit in der Zeitschrift Pacing and Clinical Electrophysiology, 2025. Verfasst von Hitoshi Mori und 10 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cardiology, Saitama Medical University, International Medical Center, Hidaka, Saitama, Japan. Hauptthemen laut MEDLINE: Pulmonary Veins, Catheter Ablation, Atrial Fibrillation, Cardiac Pacing, Artificial. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 22 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0).
Abstract (Originalsprache)
Background
The IntellaNav MiFi OI catheter (MiFi) is equipped with a sensor for local impedance (LI) monitoring and three mini‐electrodes. In this study, we investigated the target LI values for a successful pulmonary vein isolation (PVI) under the pacing and ablation technique using the MiFi catheter.
Methods
Twenty‐seven patients underwent PVI using the MiFi catheter under mini electrode pacing from the MiFi catheter. The local impedance (LI) changes, generator impedance (GI) changes, and the time to capture loss were evaluated.
Results
First‐pass isolations were obtained in 15 patients (57.7 %) for right PVs and in 22 patients (84.6 %) for left PVs. At gap sites, the impedance decrease was smaller than at non‐gap sites (non‐gap sites vs. gap sites; LI drop, 23.2 [±10.3] vs. 15.6 [±7.7] Ω, p < 0.0001; GI drop, 4.8 [±4.1] vs. 2.7 [3.9] Ω, p = 0.0026; %LI drop, −19.3 [±7.4] vs. −13.1 [±6.1] %, p < 0.0001; % GI drop, −5.1 [±4.2] vs. −2.9 [±4.2] %, p = 0.0020), suggesting that changes in impedance could be useful for predicting gaps. The cutoff values for predicting no gaps were identified as 15.0 Ω for the LI drop and −13.74% for the %LI drop.
Conclusion
The LI showed greater changes than the GI and was also useful for predicting gaps. The cutoff values of 15.0 Ω for the LI drop and −13.74% for the %LI drop could predict conduction gaps. Under the monitoring of the LI, the pacing and ablation technique proved useful for PVI, even though the MiFi catheter does not have a CF sensor or ablation indices.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Cardiology, Saitama Medical University, International Medical Center, Hidaka, Saitama, Japan.
- Department of Cardiovascular Medicine, Kitasato University School of Medicine, Sagamihara, Kanagawa, Japan.
ORCID-Profile: Hitoshi Mori, Hidehira Fukaya, Daisuke Kawano, Yoshifumi Ikeda
Zitieren
APA
Mori, H., Fukaya, H., Matsumoto, K., Narita, M., Naganuma, T., Sasaki, W., Tanaka, N., Kawano, D., Ikeda, Y., Matsumoto, K., & Kato, R. (2025). Pacing and Ablation Technique Using Microelectrode for Pulmonary Vein Isolation Using a Local Impedance‐Guided Catheter. Pacing and Clinical Electrophysiology. https://doi.org/10.1111/pace.15144
Vancouver
Mori H, Fukaya H, Matsumoto K, Narita M, Naganuma T, Sasaki W, et al. Pacing and Ablation Technique Using Microelectrode for Pulmonary Vein Isolation Using a Local Impedance‐Guided Catheter. Pacing and Clinical Electrophysiology. 2025. doi:10.1111/pace.15144
BibTeX
@article{Mori2025,
author = {Mori, Hitoshi and Fukaya, Hidehira and Matsumoto, Kazuhisa and Narita, Masataka and Naganuma, Tsukasa and Sasaki, Wataru and Tanaka, Naomichi and Kawano, Daisuke and Ikeda, Yoshifumi and Matsumoto, Kazuo and Kato, Ritsushi},
title = {Pacing and Ablation Technique Using Microelectrode for Pulmonary Vein Isolation Using a Local Impedance‐Guided Catheter},
journal = {Pacing and Clinical Electrophysiology},
year = {2025},
doi = {10.1111/pace.15144}
}
EUVIMED