Mediastinal Air Entrapment During Extravascular ICD Implantation: A Preventable Cause of Elevated Shock Impedance
Journal of Cardiovascular Electrophysiology · 2026-09 · 37 (9) · 2116-2118
DOI: 10.1111/jce.70426 ↗Kurz gefasst
Fallbericht in der Zeitschrift Journal of Cardiovascular Electrophysiology, online erschienen am 30.06.2026. Verfasst von Robert N Kerley und 1 weiteren Autor:innen. Beteiligte Einrichtung: Cardiac Arrhythmia Service, Brigham and Women's Hospital, Boston, Massachusetts, USA. Hauptthemen laut MEDLINE: Air, Defibrillators, Implantable, Electric Countershock, Mediastinum. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 5 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar.
Abstract (Originalsprache)
Introduction
Elevated shock impedance during extravascular implantable cardioverter‐defibrillator (EV‐ICD) implantation may preclude defibrillation threshold (DFT) testing despite appropriate retrosternal lead position. We describe suspected mediastinal air entrapment as a reversible mechanism and a simple preventive strategy. Methods and Results During early EV‐ICD implantation experience, two patients developed shock impedance > 200 Ω immediately after retrosternal lead placement, with associated sensing abnormalities. Lead repositioning did not normalize impedance, no pneumothorax was identified, and DFT testing was deferred. Shock impedance normalized spontaneously within 1–2 days, permitting successful DFT testing without lead revision. Transient mediastinal air entrapment along the defibrillation vector was suspected. A procedural modification was adopted in which 5–10 cm H 2 O positive end‐expiratory pressure was applied immediately before retrosternal introducer removal, followed by prompt EV‐ICD lead advancement. Among five consecutive subsequent implantations using this approach, initial shock impedance remained within the expected range, permitting immediate DFT testing without lead repositioning or delay.
Conclusion
Transient mediastinal air entrapment may cause elevated shock impedance during EV‐ICD implantation. Positive end‐expiratory pressure during introducer removal may mitigate this complication and facilitate reliable intraoperative DFT testing.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Cardiac Arrhythmia Service, Brigham and Women's Hospital, Boston, Massachusetts, USA.
ORCID-Profile: Robert N Kerley, Sunil Kapur
Zitieren
APA
Kerley, R. N., & Kapur, S. (2026). Mediastinal Air Entrapment During Extravascular ICD Implantation: A Preventable Cause of Elevated Shock Impedance. Journal of Cardiovascular Electrophysiology, 37(9), 2116-2118. https://doi.org/10.1111/jce.70426
Vancouver
Kerley RN, Kapur S. Mediastinal Air Entrapment During Extravascular ICD Implantation: A Preventable Cause of Elevated Shock Impedance. Journal of Cardiovascular Electrophysiology. 2026;37(9):2116-2118. doi:10.1111/jce.70426
BibTeX
@article{Kerley2026,
author = {Kerley, Robert N and Kapur, Sunil},
title = {Mediastinal Air Entrapment During Extravascular ICD Implantation: A Preventable Cause of Elevated Shock Impedance},
journal = {Journal of Cardiovascular Electrophysiology},
year = {2026},
volume = {37},
number = {9},
pages = {2116--2118},
doi = {10.1111/jce.70426}
}
EUVIMED