Impact of Pulmonary Vein Morphology on the Number of Pulsed Field Ablation Applications Using Circular Multielectrode Catheters
Journal of Cardiovascular Electrophysiology · 2026
DOI: 10.1111/jce.70185 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Cardiovascular Electrophysiology, 2026. Verfasst von Takahiko Nagase und 10 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cardiology, Higashiyamato Hospital, Tokyo, Japan. Hauptthemen laut MEDLINE: Pulmonary Veins, Atrial Fibrillation, Catheter Ablation, Cardiac Catheters. Die Arbeit nennt 24 Literaturangaben.
Abstract (Originalsprache)
Introduction
Pulmonary veins (PVs) morphology may affect the number of applications for acute antral PV isolation (PVI) with pulsed field ablation (PFA) using circular multielectrode catheters (PulseSelect, Medtronic) for atrial fibrillation (AF). We investigated the relationship between PV morphology and the number of PFA applications for antral PVI.
Methods
Forty‐five patients (180 PVs) with AF underwent initial PFA. PV morphology was assessed by preprocedural computed tomography. The minimal PFA protocol was defined as four ostial and four antral applications per PV. Antral PVI was confirmed after the minimal PFA protocol by multielectrode mapping catheters on 3‐dimensional electroanatomic mapping systems. PV morphology was compared between PVs with and without antral PVI by the minimal PFA protocol.
Results
Antral PVI by the minimal PFA protocol was obtained in 138 (77%) PVs. PVs without antral PVI by the minimal PFA protocol revealed larger long axis diameters (23.1 ± 5.4 vs. 20.1 ± 4.3 mm, p < 0.001), larger short axis diameters (19.0 ± 5.5 vs. 16.3 ± 4.0 mm, p = 0.005), and larger PV areas (PVAs) (362.7 ± 178.4 vs. 265.9 ± 115.2 mm 2 , p = 0.010). PVA was the only independent predictor among the clinical characteristics for antral PVI after the minimal PFA protocol (adjusted odds ratio per 10 mm 2 0.867; 95% confidence interval 0.817–0.920; p < 0.001).
Conclusions
Antral PVI was not accomplished in 23% of PVs by the minimal PFA protocol using circular multielectrode catheters. Additional PFA should be considered for larger PVs.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Cardiology, Higashiyamato Hospital, Tokyo, Japan.
Zitieren
APA
Nagase, T., Tashiro, H., Himeno, M., Hayakawa, F., Karasawa, Y., Yoshino, C., Yoshida, Y., Tanaka, T., Ishino, M., Kato, R., & Kuwada, M. (2026). Impact of Pulmonary Vein Morphology on the Number of Pulsed Field Ablation Applications Using Circular Multielectrode Catheters. Journal of Cardiovascular Electrophysiology. https://doi.org/10.1111/jce.70185
Vancouver
Nagase T, Tashiro H, Himeno M, Hayakawa F, Karasawa Y, Yoshino C, et al. Impact of Pulmonary Vein Morphology on the Number of Pulsed Field Ablation Applications Using Circular Multielectrode Catheters. Journal of Cardiovascular Electrophysiology. 2026. doi:10.1111/jce.70185
BibTeX
@article{Nagase2026,
author = {Nagase, Takahiko and Tashiro, Haruwo and Himeno, Masafumi and Hayakawa, Fumiaki and Karasawa, Yuka and Yoshino, Chiyo and Yoshida, Yoshinori and Tanaka, Takahisa and Ishino, Mitsunori and Kato, Ryuichi and Kuwada, Masao},
title = {Impact of Pulmonary Vein Morphology on the Number of Pulsed Field Ablation Applications Using Circular Multielectrode Catheters},
journal = {Journal of Cardiovascular Electrophysiology},
year = {2026},
doi = {10.1111/jce.70185}
}
EUVIMED