Difference in Microembolic Signals Between Flex‐Tip Versus Non‐Flex‐Tip Type Catheters During High‐Power Short‐Duration Ablation

Yoshiaki Mizutani iD, Satoshi Yanagisawa iD, Mizuki Ichikawa, Keisuke Nishio, Hiroya Sakai, Daishi Nonokawa, Yuichiro Makino, Hitomi Suzuki, Hitoshi Ichimiya, Yasuhiro Uchida, Junji Watanabe, Masaaki Kanashiro und 6 weitere

Journal of Cardiovascular Electrophysiology · 2025

DOI: 10.1111/jce.70074 ↗

Kurz gefasst

Vergleichsstudie in der Zeitschrift Journal of Cardiovascular Electrophysiology, 2025. Verfasst von Yoshiaki Mizutani und 17 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cardiology, Yokkaichi Municipal Hospital, Yokkaichi, Japan. Hauptthemen laut MEDLINE: Catheter Ablation, Atrial Fibrillation, Cardiac Catheters, Pulmonary Veins. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 24 Literaturangaben. Gefördert von: This study was supported by a Grant-in-Aid for Scientific Research (C) (23K07549) and Suzuken Memorial Foundation 23-095..

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Introduction

A novel flex‐tip ablation catheter has a distal tip with a unique design, especially for optimized irrigation flow against catheter contact. This study aimed to evaluate the microembolic signals (MESs) during high‐power short‐duration (HPSD) ablation between flex‐tip and non‐flex‐tip catheters. Method and Results We retrospectively evaluated 70 patients (39 and 31 with flex‐tip and non‐flex‐tip catheters, respectively) who underwent first‐time HPSD radiofrequency ablation with a 50 W‐power setting for atrial fibrillation at two institutions. All patients were clinically assessed for MESs using a novel carotid ultrasound‐Doppler system throughout pulmonary vein isolation. The numbers of MESs were compared between the two types of catheters. In an experimental study, porcine heart pieces were ablated using two catheters to observe MESs using the same ultrasound evaluation. In the clinical study, 4913 points (2932 and 1981 points in the flex‐tip and non‐flex‐tip groups, respectively) were analyzed. The numbers of MESs, MESs per point, and MESs per second were significantly lower in the flex‐tip group than in the non‐flex‐tip group (123 ± 171 vs. 332 ± 256, p < 0.001; 1.8 ± 2.7 vs. 5.9 ± 4.3, p < 0.001; and 0.1 ± 0.1 vs. 0.2 ± 0.2, p < 0.001, respectively). After propensity score matching analysis in the clinical study with 19 patients in each group, the number of MESs was significantly lower in the flex‐tip group than in the non‐flex‐tip group (152 ± 228 vs. 378 ± 300, p = 0.013), and the difference was most evident in bubble‐type MESs (24 [16–96] vs. 185 [79–421], p = 0.023). In the experimental study with eight attempt ablations in each catheter, the number of MESs was significantly lower in the flex‐tip group than in the non‐flex‐tip group (26 ± 14 vs. 58 ± 17, p = 0.002).

Conclusions

Flex‐tip catheter ablation significantly reduced MES detections, particularly in the bubble‐type MESs, compared with the non‐flex‐tip catheter ablation at 50 W. This may indicate safe HPSD ablation with the flex‐tip type catheter.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Catheter AblationAtrial FibrillationCardiac CathetersPulmonary VeinsHumansRetrospective StudiesMaleFemaleMiddle AgedAnimalsTime FactorsTreatment OutcomeEquipment DesignAgedSwineRisk FactorsAction Potentials

Stichwörter der Autor:innen

flex‐tip catheterhigh‐power short‐durationinternal carotid artery ultrasoundmicroembolic signalspulmonary vein isolation

Autor:innen und Institutionen

  • Department of Cardiology, Yokkaichi Municipal Hospital, Yokkaichi, Japan.
  • Department of Cardiology, Nagoya University Graduate School of Medicine, Nagoya, Japan.
  • Department of Clinical Laboratory, Yokkaichi Municipal Hospital, Yokkaichi, Japan.
  • Department of Clinical Engineering, Nagoya University Hospital, Nagoya, Japan.

ORCID-Profile: Yoshiaki Mizutani, Satoshi Yanagisawa, Yasuya Inden

Förderung

  • This study was supported by a Grant-in-Aid for Scientific Research (C) (23K07549) and Suzuken Memorial Foundation 23-095.

Zitieren

APA

Mizutani, Y., Yanagisawa, S., Ichikawa, M., Nishio, K., Sakai, H., Nonokawa, D., Makino, Y., Suzuki, H., Ichimiya, H., Uchida, Y., Watanabe, J., Kanashiro, M., Iwawaki, T., Kondo, S., Yuki Sato, Shibata, R., Inden, Y., & Murohara, T. (2025). Difference in Microembolic Signals Between Flex‐Tip Versus Non‐Flex‐Tip Type Catheters During High‐Power Short‐Duration Ablation. Journal of Cardiovascular Electrophysiology. https://doi.org/10.1111/jce.70074

Vancouver

Mizutani Y, Yanagisawa S, Ichikawa M, Nishio K, Sakai H, Nonokawa D, et al. Difference in Microembolic Signals Between Flex‐Tip Versus Non‐Flex‐Tip Type Catheters During High‐Power Short‐Duration Ablation. Journal of Cardiovascular Electrophysiology. 2025. doi:10.1111/jce.70074

BibTeX
@article{Mizutani2025,
  author = {Mizutani, Yoshiaki and Yanagisawa, Satoshi and Ichikawa, Mizuki and Nishio, Keisuke and Sakai, Hiroya and Nonokawa, Daishi and Makino, Yuichiro and Suzuki, Hitomi and Ichimiya, Hitoshi and Uchida, Yasuhiro and Watanabe, Junji and Kanashiro, Masaaki and Iwawaki, Tomoya and Kondo, Shun and Yuki Sato and Shibata, Rei and Inden, Yasuya and Murohara, Toyoaki},
  title = {Difference in Microembolic Signals Between Flex‐Tip Versus Non‐Flex‐Tip Type Catheters During High‐Power Short‐Duration Ablation},
  journal = {Journal of Cardiovascular Electrophysiology},
  year = {2025},
  doi = {10.1111/jce.70074}
}
Kostenlos starten →