High‐Power Short‐Duration Temperature‐Controlled Radiofrequency Ablation for the Treatment of Outflow Tract Ventricular Arrhythmias
Journal of Cardiovascular Electrophysiology · 2025
DOI: 10.1111/jce.16636 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Cardiovascular Electrophysiology, 2025. Verfasst von Dionyssios Leftheriotis und 6 weiteren Autor:innen. Beteiligte Einrichtung: 2nd Department of Cardiology, Medical School, National and Kapodistrian University of Athens, University General Hospital ATTIKON, Athens, Greece. Hauptthemen laut MEDLINE: Catheter Ablation, Ventricular Premature Complexes, Tachycardia, Ventricular. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 31 Literaturangaben.
Abstract (Originalsprache)
Introduction
The tissue temperature‐controlled DiamondTemp ablation (DTA) catheter has been mainly used for atrial fibrillation ablation. We report our initial experience in using this catheter for the treatment of outflow premature ventricular contractions (PVCs) or repetitive non‐sustained monomorphic ventricular tachycardias (VTs).
Methods
Twenty patients were studied: 10 with right ventricular outflow tract PVCs/VTs, eight with PVCs/VTs from the aortic sinus cusps, and two with left ventricular outflow tract PVCs. The DTA parameters were adjusted to a target‐temperature of 60°C with 50 W power. Ablation success was defined as the absence of clinical VT/PVCs at the end of the procedure and during the next 48 h of continuous rhythm monitoring without antiarrhythmic drugs. All patients were reevaluated 2 months after the ablation.
Results
The average ablation duration was 262 (145–690) seconds and the average tissue temperature (T) was 50.4 ± 2.9°C. A mean number of 11 ± 6 lesions per case were applied. The average time of T > 50°C was 18 (12–20) seconds, that is 70% of the ablation time, and an average relative impedance drop of 13.8 ± 2.0% was recorded. No serious complications occurred, and no VT recurrence was observed until discharge and during the short follow‐up period.
Conclusion
This initial evaluation suggests that the DTA system reached the adjusted temperature and power targets, and it can be effectively and safely applied for the treatment of outflow tract PVCs/VTs. This performance should be further evaluated by long‐term randomized controlled trials in comparison to conventional ablation catheters.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- 2nd Department of Cardiology, Medical School, National and Kapodistrian University of Athens, University General Hospital ATTIKON, Athens, Greece.
- Department of Nursing, University of West Attica, Athens, Greece.
ORCID-Profile: Konstantinos A Papathanasiou
Zitieren
APA
Leftheriotis, D., Flevari, P., Papathanasiou, K. A., Karamanolis, E., Gkatzia, A., Polikandrioti, M., & Filippatos, G. (2025). High‐Power Short‐Duration Temperature‐Controlled Radiofrequency Ablation for the Treatment of Outflow Tract Ventricular Arrhythmias. Journal of Cardiovascular Electrophysiology. https://doi.org/10.1111/jce.16636
Vancouver
Leftheriotis D, Flevari P, Papathanasiou KA, Karamanolis E, Gkatzia A, Polikandrioti M, et al. High‐Power Short‐Duration Temperature‐Controlled Radiofrequency Ablation for the Treatment of Outflow Tract Ventricular Arrhythmias. Journal of Cardiovascular Electrophysiology. 2025. doi:10.1111/jce.16636
BibTeX
@article{Leftheriotis2025,
author = {Leftheriotis, Dionyssios and Flevari, Panagiota and Papathanasiou, Konstantinos A and Karamanolis, Efstratios and Gkatzia, Alexandra and Polikandrioti, Maria and Filippatos, Gerasimos},
title = {High‐Power Short‐Duration Temperature‐Controlled Radiofrequency Ablation for the Treatment of Outflow Tract Ventricular Arrhythmias},
journal = {Journal of Cardiovascular Electrophysiology},
year = {2025},
doi = {10.1111/jce.16636}
}
EUVIMED