Ventricular tachycardia ablation with pentaspline pulsed field technology in two patients with ischemic cardiomyopathy

Gaetano Fassini, Elio Zito, Lorenzo Bianchini, Fabrizio Tundo, Claudio Tondo, Marco Schiavone iD

Journal of Cardiovascular Electrophysiology · 2024

DOI: 10.1111/jce.16418 ↗

Kurz gefasst

Fallbericht in der Zeitschrift Journal of Cardiovascular Electrophysiology, 2024. Verfasst von Gaetano Fassini und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Clinical Electrophysiology & Cardiac Pacing, Centro Cardiologico Monzino, IRCCS, Milan, Italy. Hauptthemen laut MEDLINE: Tachycardia, Ventricular, Myocardial Ischemia, Catheter Ablation, Cardiomyopathies. Die Arbeit wurde bisher 10-mal zitiert (Crossref) und nennt 13 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Introduction

Due to its unique features, pulsed field ablation (PFA) could potentially overcome some limitations of current radiofrequency (RF) ventricular tachycardia (VT) ablation. However, data on the use of PFA in this setting are currently scarce.

Methods

Two patients with ischemic cardiomyopathy and previously failed RF VT ablations were treated with PFA.

Results

A total of 18 bipolar applications (case1) and seven bipolar applications (case2) were delivered to the infero‐lateral and infero‐septal areas (case1) and to the apical lateral left ventricular (LV) wall (case2), placing the catheter adjacent to the LV wall in the flower configuration. A rapid cessation of VT and restoration of sinus rhythm were observed during PFA delivery in both cases. Further applications were delivered to achieve complete elimination of late potentials. In case 1, during the in‐hospital stay, ECG monitoring did not show VT recurrences. Six‐month follow‐up was uneventful, with no VT recurrences at ICD interrogation. In case 2, due to postdischarge VT recurrences, a second RF procedure was scheduled 1 month later. The voltage map performed in sinus rhythm showed a low‐voltage zone located at the anterolateral wall, near the previous ablation site. Numerous late potentials were recorded. At the 6‐month follow‐up, no further VT recurrences were documented after RF redo ablation.

Conclusion

While the speed of application and potential transmural effect can facilitate the ablation of large diseased endocardial areas, early loss of contact due to difficult pentaspline catheter manipulation in the LV could lead to insufficient contact force and, consequently, inadequate energy penetration.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Tachycardia, VentricularMyocardial IschemiaCatheter AblationCardiomyopathiesHeart RateHumansMaleAgedTreatment OutcomeAction PotentialsMiddle AgedElectrophysiologic Techniques, CardiacFemaleTime Factors

Stichwörter der Autor:innen

ischemic cardiomyopathypulsed field ablationradiofrequencyscar‐relatedventricular tachycardia

Autor:innen und Institutionen

  • Department of Clinical Electrophysiology & Cardiac Pacing, Centro Cardiologico Monzino, IRCCS, Milan, Italy.
  • Department of Biomedical, Surgical and Dental Sciences, University of Milan, Milan, Italy.
  • Department of Systems Medicine, University of Rome Tor Vergata, Rome, Italy.

ORCID-Profile: Marco Schiavone

Zitieren

APA

Fassini, G., Zito, E., Bianchini, L., Tundo, F., Tondo, C., & Schiavone, M. (2024). Ventricular tachycardia ablation with pentaspline pulsed field technology in two patients with ischemic cardiomyopathy. Journal of Cardiovascular Electrophysiology. https://doi.org/10.1111/jce.16418

Vancouver

Fassini G, Zito E, Bianchini L, Tundo F, Tondo C, Schiavone M. Ventricular tachycardia ablation with pentaspline pulsed field technology in two patients with ischemic cardiomyopathy. Journal of Cardiovascular Electrophysiology. 2024. doi:10.1111/jce.16418

BibTeX
@article{Fassini2024,
  author = {Fassini, Gaetano and Zito, Elio and Bianchini, Lorenzo and Tundo, Fabrizio and Tondo, Claudio and Schiavone, Marco},
  title = {Ventricular tachycardia ablation with pentaspline pulsed field technology in two patients with ischemic cardiomyopathy},
  journal = {Journal of Cardiovascular Electrophysiology},
  year = {2024},
  doi = {10.1111/jce.16418}
}
Kostenlos starten →