EUVIMEDEUVIMED entdecken
VergleichsstudieOpen Access · CC BY-NC-ND 4.0

Overestimation of Automated Platelet Counts by Blast Fragments in Acute Hypercellular Leukemias: A Retrospective Study Comparing Impedance, Optical ( PLT ‐O), Fluorescent ( PLT ‐F), and CD41 / CD61 Flow Cytometry Methods

Victor Bobée iD, Romain Ravel-Chapuis, Maïssa Souissi, Catherine Boutet, Dany Bigot, Cédric Paquin, Jorel Gruchet, Virginie Barbay, Vincent Camus, Elsa Bera, Sylvie Daliphard

International Journal of Laboratory Hematology · 2025

DOI: 10.1111/ijlh.14464 ↗

Kurz gefasst

Vergleichsstudie in der Zeitschrift International Journal of Laboratory Hematology, 2025. Verfasst von Victor Bobée und 10 weiteren Autor:innen. Beteiligte Einrichtung: Department of Biological Hematology, Rouen University Hospital, Rouen, France. Hauptthemen laut MEDLINE: Flow Cytometry, Leukemia, Leukemia, Myeloid, Acute. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 41 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Introduction

Acute leukemias with hyperleukocytosis (> 100 × 10 9 /L) can produce cytoplasmic blast fragments that interfere with platelet counts, notably in impedance‐based methods, potentially masking severe thrombocytopenia and increasing hemorrhagic risk. While fluorescent platelet counting (PLT‐F) is promoted as platelet‐specific, its accuracy in the presence of blast fragments remains uncertain.

Methods

This retrospective study analyzed 269 blood samples from 87 patients with hypercellular acute leukemia. Blast fragments were identified on blood smears. Platelet counts by impedance were compared to optical (PLT‐O) and fluorescent (PLT‐F) methods when available. Flow cytometry (FC) quantification by CD41+/CD61+ staining was performed in selected cases.

Results

Blast fragments were observed in 26% of cases, predominantly in AML‐M5 and AML‐M1 subtypes. In the absence of blast fragments, PLT‐I, PLT‐O, and PLT‐F showed comparable results. However, in samples with blast fragments, PLT‐I frequently overestimated platelet counts compared to PLT‐O and PLT‐F. PLT‐F counts were closer to FC quantification but still overestimated platelet numbers in 6 of 16 samples with FC results, particularly in cases of severe leukocytosis. Notably, PLT‐F failed to trigger abnormal scattergram flags in all but one of the discrepant cases. PLT‐O provided results comparable to PLT‐F in most cases with blast fragments but also demonstrated limitations in select cases.

Conclusion

PLT‐O and PLT‐F methods both have limitations in hypercellular acute leukemias with blast fragments. FC remains the most reliable approach when blast fragments are present. Routine blood smear evaluations are essential for detecting interferences and ensuring accurate thrombocytopenia assessment in these high‐risk patients.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Flow CytometryLeukemiaLeukemia, Myeloid, AcuteHumansRetrospective StudiesPlatelet CountMaleFemaleMiddle AgedAdultIntegrin beta3AgedPlatelet Membrane Glycoprotein IIbBlood PlateletsAcute DiseaseElectric Impedance

Stichwörter der Autor:innen

Sysmexacute leukemiasflow cytometrylaboratory practiceplateletsthrombocytopenia

Substanzen

Integrin beta3Platelet Membrane Glycoprotein IIbITGB3 protein, human

Autor:innen und Institutionen

  • Department of Biological Hematology, Rouen University Hospital, Rouen, France.
  • Department of Biology, Centre Henri Becquerel, Rouen, France.
  • Department of Hematology, Centre Henri Becquerel, Rouen, France.

ORCID-Profile: Victor Bobée

Zitieren

APA

Bobée, V., Ravel-Chapuis, R., Souissi, M., Boutet, C., Bigot, D., Paquin, C., Gruchet, J., Barbay, V., Camus, V., Bera, E., & Daliphard, S. (2025). Overestimation of Automated Platelet Counts by Blast Fragments in Acute Hypercellular Leukemias: A Retrospective Study Comparing Impedance, Optical ( PLT ‐O), Fluorescent ( PLT ‐F), and CD41 / CD61 Flow Cytometry Methods. International Journal of Laboratory Hematology. https://doi.org/10.1111/ijlh.14464

Vancouver

Bobée V, Ravel-Chapuis R, Souissi M, Boutet C, Bigot D, Paquin C, et al. Overestimation of Automated Platelet Counts by Blast Fragments in Acute Hypercellular Leukemias: A Retrospective Study Comparing Impedance, Optical ( PLT ‐O), Fluorescent ( PLT ‐F), and CD41 / CD61 Flow Cytometry Methods. International Journal of Laboratory Hematology. 2025. doi:10.1111/ijlh.14464

BibTeX
@article{Bobe2025,
  author = {Bobée, Victor and Ravel-Chapuis, Romain and Souissi, Maïssa and Boutet, Catherine and Bigot, Dany and Paquin, Cédric and Gruchet, Jorel and Barbay, Virginie and Camus, Vincent and Bera, Elsa and Daliphard, Sylvie},
  title = {Overestimation of Automated Platelet Counts by Blast Fragments in Acute Hypercellular Leukemias: A Retrospective Study Comparing Impedance, Optical ( PLT ‐O), Fluorescent ( PLT ‐F), and CD41 / CD61 Flow Cytometry Methods},
  journal = {International Journal of Laboratory Hematology},
  year = {2025},
  doi = {10.1111/ijlh.14464}
}
Kostenlos starten →