Non‐Invasive Intracranial Pressure Waveform Assessment as a Screening Test for Abnormal Lumbar Puncture Results
European Journal of Neurology · 2026
DOI: 10.1111/ene.70776 ↗Kurz gefasst
Fachartikel in der Zeitschrift European Journal of Neurology, 2026. Verfasst von Gabriel Taricani Kubota und 11 weiteren Autor:innen. Beteiligte Einrichtung: Department of Neurology, University of Sao Paulo, Sao Paulo, Brazil. Hauptthemen laut MEDLINE: Spinal Puncture, Intracranial Pressure, Intracranial Hypertension. Die Arbeit nennt 37 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0). Gefördert von: LIM62 - Pain Center of the Department of Neurology at the University of São Paulo, Brain4Care Brazil.
Abstract (Originalsprache)
Background
Lumbar puncture (LP) is essential for detecting intracranial hypertension (IH) but is often non‐diagnostic and not without complications. A skull extensometer‐based non‐invasive intracranial pressure waveform assessment (NIWA) has shown promise in detecting IH in neurocritical settings. This study investigated its feasibility and accuracy among outpatient neurological individuals.
Methods
This cross‐sectional study enrolled adults referred to routine outpatient LP at a tertiary‐care hospital. Subjects underwent baseline supine NIWA, then were sequentially divided (2:1:1 ratio) to: LP then follow‐up supine NIWA; follow‐up NIWA while sitting then LP; or follow‐up NIWA during Valsalva maneuver then LP. Several waveform parameters and their variation between baseline and follow‐up assessments were compared with LP results. IH was defined as LP opening pressure > 25cmH 2 O.
Results
The study included 130 subjects, and 28 (21.5%) presented with IH. Dicrotic peak slope < 0.143 predicted IH with 89.5% sensitivity and 29.7% specificity (AUROC = 0.680, 95% CI 0.541–0.819). Equations derived from multivariate logistic regression models including mostly latency‐related parameters of the entire waveform and its percussion and dicrotic peaks predicted cerebrospinal fluid (CSF) chemocytology changes with AUROC = 0.661 (95% CI 0.542–0.780) and overall abnormal LP results (IH and/or altered CSF chemocytology) with 96.2% sensitivity and 24.1% specificity (AUROC = 0.696, 95% CI 0.572–0.820). Parameter changes after LP, during Valsalva maneuver, and sitting position did not improve diagnostic accuracy.
Conclusions
NIWA was feasible and detected IH and CSF changes with high sensitivity among real‐world neurological outpatients. While LP remains pivotal to many cases, pre‐LP screening with NIWA may enhance diagnostic yield and mitigate complications and healthcare‐related costs.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Neurology, University of Sao Paulo, Sao Paulo, Brazil.
- Pain Treatment Center, Sao Paulo State Cancer Institute, University of Sao Paulo Clinics Hospital, Sao Paulo, Brazil.
- Brain4Care Brazil, São Carlos, Brazil.
- Cerebrospinal Fluid Sector, LIM-15, Central Laboratory Division, University of Sao Paulo Clinics Hospital, Sao Paulo, Brazil.
- Service of Interdisciplinary Neuromodulation (SIN), Department and Institute of Psychiatry, University of São Paulo, Sao Paulo, Brazil.
- Physical Medicine and Rehabilitation Division, Orthopaedics and Traumatology Institute, University of Sao Paulo Clinics Hospital, Sao Paulo, Brazil.
ORCID-Profile: Gabriel Taricani Kubota, Cintya Yukie Hayashi, Valquíria Aparecida da Silva, Eberval Gadelha Figueiredo, Lin Tchia Yeng, Manoel Jacobsen Teixeira, Daniel Ciampi de Andrade
Förderung
- LIM62 - Pain Center of the Department of Neurology at the University of São Paulo
- Brain4Care Brazil
Zitieren
APA
Kubota, G. T., da Fonseca, A. S., Hayashi, C. Y., de Oliveira Pereira, R. G., Gonçalves, T., Gomes, H. R., da Silva, V. A., Fortini, I., Figueiredo, E. G., Yeng, L. T., Teixeira, M. J., & de Andrade, D. C. (2026). Non‐Invasive Intracranial Pressure Waveform Assessment as a Screening Test for Abnormal Lumbar Puncture Results. European Journal of Neurology. https://doi.org/10.1111/ene.70776
Vancouver
Kubota GT, da Fonseca AS, Hayashi CY, de Oliveira Pereira RG, Gonçalves T, Gomes HR, et al. Non‐Invasive Intracranial Pressure Waveform Assessment as a Screening Test for Abnormal Lumbar Puncture Results. European Journal of Neurology. 2026. doi:10.1111/ene.70776
BibTeX
@article{Kubota2026,
author = {Kubota, Gabriel Taricani and da Fonseca, Alessandra Santos and Hayashi, Cintya Yukie and de Oliveira Pereira, Rebecca Gabriela and Gonçalves, Thauan and Gomes, Helio Rodrigues and da Silva, Valquíria Aparecida and Fortini, Ida and Figueiredo, Eberval Gadelha and Yeng, Lin Tchia and Teixeira, Manoel Jacobsen and de Andrade, Daniel Ciampi},
title = {Non‐Invasive Intracranial Pressure Waveform Assessment as a Screening Test for Abnormal Lumbar Puncture Results},
journal = {European Journal of Neurology},
year = {2026},
doi = {10.1111/ene.70776}
}
EUVIMED