Validation of a smart shirt for heart rate variability measurements at rest and during exercise

Peter Schams, Boris Feodoroff, Jonas Zacher, Angi Eibl, Ingo Froböse

Clinical Physiology and Functional Imaging · 2022-05 · 42 (3) · 190-199

DOI: 10.1111/cpf.12746 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Clinical Physiology and Functional Imaging, online erschienen am 28.03.2022. Verfasst von Peter Schams und 4 weiteren Autor:innen. Beteiligte Einrichtung: Institute of Movement Therapy and Movement-oriented Prevention and Rehabilitation Sciences, German Sport University, Cologne, Germany. Hauptthemen laut MEDLINE: Electrocardiography, Exercise. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 46 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: Bundesministerium für Wirtschaft und Energie.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Heart rate variability (HRV) monitoring is a promising option to estimate the autonomic nervous system regulation responding to exercise. Textiles with embedded sensors recording heartbeat intervals are a simple tool for data collection. The so‐called smart shirts offer comfort for daily use and are managed easily. Their measurement accuracy for HRV calculation at rest is promising, but remains questionable during exercise. Therefore, the present study validated the Ambiotex smart shirt using HRV indices (root mean square of successive differences, rel. HF power [high‐frequency power percentage of total power] and rel. LF [low‐frequency power percentage of total power] power) during exercise. Eighty‐three healthy participants (31 ± 6 years; 39 females, 44 males) completed an incremental exercise test on a bicycle ergometer wearing the smart shirt and an electrocardiogram simultaneously. We compared HRV indices of segments at rest (5 min), at warm‐up (3 min) and twice at the exercise test (each 5 min). At rest and at warm‐up, we observed excellent linear relationship ( r > 0.96; R 2 ​​​​​ > 0.94), excellent relative reliability (intraclass correlation coefficient ≥ 0.98; α ≥ 0.98) and acceptable agreement (bias < 10%). During the exercise test, measurement accuracy declined with increasing intensity but remained high (>0.8), although results for partial HRV indices were insufficient. In addition, percentage bias was unacceptable during an exercise test. However, the findings support the validity of the smart shirt for measuring HRV, especially at rest and at warm‐up. We suggest using the smart shirt for monitoring HRV indices on a daily basis, but caution should be taken in the interpretation of HRV indices obtained during moderate to vigorous exercise intensities.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

ElectrocardiographyExerciseExercise TestFemaleHeart RateHumansMaleReproducibility of Results

Stichwörter der Autor:innen

ECGHRVbicycle ergometercomparison studyheart rate monitorincremental exercise testmeasurement accuracy

Autor:innen und Institutionen

  • Institute of Movement Therapy and Movement-oriented Prevention and Rehabilitation Sciences, German Sport University, Cologne, Germany.
  • Institute of Cardiology and Sports Medicine; German Sport University, Cologne, Germany.

Förderung

  • Bundesministerium für Wirtschaft und Energie

Zitieren

APA

Schams, P., Feodoroff, B., Zacher, J., Eibl, A., & Froböse, I. (2022). Validation of a smart shirt for heart rate variability measurements at rest and during exercise. Clinical Physiology and Functional Imaging, 42(3), 190-199. https://doi.org/10.1111/cpf.12746

Vancouver

Schams P, Feodoroff B, Zacher J, Eibl A, Froböse I. Validation of a smart shirt for heart rate variability measurements at rest and during exercise. Clinical Physiology and Functional Imaging. 2022;42(3):190-199. doi:10.1111/cpf.12746

BibTeX
@article{Schams2022,
  author = {Schams, Peter and Feodoroff, Boris and Zacher, Jonas and Eibl, Angi and Froböse, Ingo},
  title = {Validation of a smart shirt for heart rate variability measurements at rest and during exercise},
  journal = {Clinical Physiology and Functional Imaging},
  year = {2022},
  volume = {42},
  number = {3},
  pages = {190--199},
  doi = {10.1111/cpf.12746}
}