Accuracy of a Novel Intraoral Photogrammetry Technique for Complete‐Arch Implant Impressions: An In Vitro Study
Clinical Oral Implants Research · 2025
DOI: 10.1111/clr.14445 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift Clinical Oral Implants Research, 2025. Verfasst von Xiao-Jiao Fu und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Oral and Maxillofacial Implantology, Shanghai PerioImplant Innovation Center, Shanghai Ninth People's Hospital, Shanghai Jiao Tong Univ…. Hauptthemen laut MEDLINE: Photogrammetry, Dental Impression Technique, Dental Implants. Die Arbeit wurde bisher 24-mal zitiert (Crossref) und nennt 33 Literaturangaben. Gefördert von: the Project of Biobank, the Biomaterials and Regenerative Medicine Institute Cooperative Research Project by Shanghai JiaoTong University School of Medicine, the Shanghai Rising Star Project.
Abstract (Originalsprache)
Objectives
To compare the accuracy of complete‐arch implant impressions using four digital techniques: extraoral photogrammetry ( EPG ), intraoral scanning with original scan body ( IOS ), intraoral scanning with prefabricated aids ( IOSA ), and intraoral photogrammetry ( IPG ). Materials and Methods One edentulous maxillary master model with six parallel abutment analogs was scanned by a laboratory scanner as the reference scan. EPG , IOS , IOSA , and IPG were used to scan the master model with respective scan bodies as test scans. Ten scanning data were gained for each technique. Trueness and precision of root mean square ( RMS ) errors were measured between the test and reference scans. Deviations in distance and angle relative to the reference scan between all pairs of abutment analogs were measured. Scanning time was recorded.
Results
IPG showed significantly best trueness in RMS errors (IPG, 26.37 ± 1.02 μm; EPG, 31.72 ± 0.59 μm; IOS, 39.93 ± 7.98 μm; IOSA, 50.60 ± 12.46 μm), (all p < 0.050). EPG showed significantly best precision in RMS errors (EPG, 2.30 ± 1.31 μm; IPG , 4.12 ± 0.87 μm; IOS , 33.38 ± 13.95 μm; IOSA , 28.35 ± 11.48 μm), (all p < 0.001). IPG demonstrated significantly best performance in distance deviation ( IPG , 28.59 ± 24.75 μm; EPG , 55.55 ± 37.82 μm; IOS , 62.97 ± 50.60 μm; IOSA , 93.37 ± 72.15 μm), (all p < 0.050). No significant differences were found regarding angle deviation ( p = 0.313). IOSA showed the longest scanning duration ( IPG : 50.30 ± 6.77 s; EPG : 57.40 ± 5.19 s; IOS : 91.10 ± 20.31 s; IOSA : 125.00 ± 18.74 s), (all p < 0.001).
Conclusions
IPG achieved comparable accuracy with EPG in complete‐arch implant digital impressions in vitro. IPG offers an efficient and straightforward workflow, making it a valuable alternative method; however, further clinical studies are needed to validate its efficacy.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Oral and Maxillofacial Implantology, Shanghai PerioImplant Innovation Center, Shanghai Ninth People's Hospital, Shanghai Jiao Tong University School of Medicine, College of Stomatology, Shanghai Jiao Tong…
- European Research Group on Periodontology, Genova, Italy.
ORCID-Profile: Xiao-Jiao Fu, Maurizio S Tonetti, Hong-Chang Lai
Förderung
- the Project of Biobank
- the Biomaterials and Regenerative Medicine Institute Cooperative Research Project by Shanghai JiaoTong University School of Medicine
- the Shanghai Rising Star Project
- the Fundamental Research Funds for the Central Universities
- Fundamental Research Funds for the Central Universities
Zitieren
APA
Fu, X. J., Cai, Z. Z., Shi, J. Y., Qiao, S. C., Tonetti, M. S., Lai, H. C., & Liu, B. L. (2025). Accuracy of a Novel Intraoral Photogrammetry Technique for Complete‐Arch Implant Impressions: An In Vitro Study. Clinical Oral Implants Research. https://doi.org/10.1111/clr.14445
Vancouver
Fu XJ, Cai ZZ, Shi JY, Qiao SC, Tonetti MS, Lai HC, et al. Accuracy of a Novel Intraoral Photogrammetry Technique for Complete‐Arch Implant Impressions: An In Vitro Study. Clinical Oral Implants Research. 2025. doi:10.1111/clr.14445
BibTeX
@article{Fu2025,
author = {Fu, Xiao-Jiao and Cai, Zheng-Zhen and Shi, Jun-Yu and Qiao, Shi-Chong and Tonetti, Maurizio S and Lai, Hong-Chang and Liu, Bei-Lei},
title = {Accuracy of a Novel Intraoral Photogrammetry Technique for Complete‐Arch Implant Impressions: An In Vitro Study},
journal = {Clinical Oral Implants Research},
year = {2025},
doi = {10.1111/clr.14445}
}
EUVIMED