EUVIMEDEUVIMED entdecken
Vergleichsstudie

Performance of point-of-care creatinine testing in the estimation of glomerular filtration rate: comparison with laboratory creatinine and cystatin C-based measurements

Ibrahima Lyra Sarr iD, Cheikh Faye, Modou Ndongo, Babacar Mbodj, Baratou Coundoul, Papa Aly Thiam Gueye, Mame Cheikh Seck, Sidy Mohamed Seck

Journal of Nephrology · 2026

DOI: 10.1093/joneph/aajag080 ↗

Kurz gefasst

Vergleichsstudie in der Zeitschrift Journal of Nephrology, 2026. Verfasst von Ibrahima Lyra Sarr und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Nephrology, Dialysis and Kidney Transplantation, Ouakam Military Hospital, Dakar, Senegal. Hauptthemen laut MEDLINE: Glomerular Filtration Rate, Cystatin C, Creatinine, Renal Insufficiency, Chronic. Die Arbeit nennt 22 Literaturangaben. Gefördert von: Centre d'Excellence African Santé et Environnement, Centre d’Excellence African Santé et Environnement.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Chronic kidney disease (CKD) is a major global public health issue, especially in low- and middle-income countries where access to laboratory testing is limited. Point-of-care tests for creatinine may offer an alternative for detection of CKD with kidney function impairment in such settings. However, its analytical and clinical validity compared to standard laboratory methods remains uncertain.

Aim

To assess the performance of a point-of-care test device for estimating glomerular filtration rate (GFR) compared to laboratory-based creatinine and cystatin C-based estimated GFR (eGFR).

Methods

This cross-sectional analytical study was conducted at Ouakam Military Hospital and included 405 adults undergoing routine kidney function testing. Each participant underwent point-of-care testing using the Nova StatSensor® device and standard laboratory tests for serum creatinine and cystatin C. The eGFR was calculated using the 2021 Chronic Kidney Disease Epidemiology Collaboration (CKD-EPI) equations. Performance was evaluated through correlation analyses, Bland-Altman plots, Deming regression, percentage of values within ±30% of the reference value (P30) accuracy, and diagnostic performance using receiver operating characteristic (ROC) curves.

Results

Point-of-care tests for creatinine showed moderate correlation with laboratory serum creatinine (r = 0.48) and respective eGFR values (r = 0.60), but consistently overestimated creatinine, leading to underestimation of eGFR (mean bias: −31.09 mL/min/1.73 m2). Correlation with cystatin C-based eGFR was weaker (r = 0.35), with a significant negative bias (−38.57 mL/min/1.73 m2). P30 accuracy was low (32.6% vs lab creatinine; 28.1% vs cystatin C). The diagnostic performance for detecting CKD stage 3 and over (eGFR < 60 mL/min/1.73 m2) demonstrated a high negative predictive value (91.3% vs 90% using cystatin C and serum creatinine, respectively, as reference).

Conclusions

While point-of-care testing shows limitations for precise GFR estimation, it demonstrates utility for community-based CKD screening with high negative predictive value.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Glomerular Filtration RateCystatin CCreatinineRenal Insufficiency, ChronicPoint-of-Care TestingHumansCross-Sectional StudiesFemaleMaleMiddle AgedAgedAdultBiomarkersReproducibility of ResultsPredictive Value of TestsKidney Function Tests

Stichwörter der Autor:innen

estimation of glomerular filtration ratelaboratory measurementsperformancepoint-of-care creatinine testing

Substanzen

Cystatin CCreatinineBiomarkers

Autor:innen und Institutionen

  • Department of Nephrology, Dialysis and Kidney Transplantation, Ouakam Military Hospital, Dakar, Senegal.
  • Department of Medical Biochemistry, Ouakam Military Hospital, Dakar, Senegal.
  • Department of Nephrology and Dialysis, Amath Dansokho Hospital, Kedougou, Senegal.
  • Department of Parasitology, Cheikh Anta Diop University, Dakar, Senegal.
  • Faculty of Health Sciences, Gaston Berger University, Saint Louis, Senegal.

ORCID-Profile: Ibrahima Lyra Sarr

Förderung

  • Centre d'Excellence African Santé et Environnement
  • Centre d’Excellence African Santé et Environnement

Zitieren

APA

Sarr, I. L., Faye, C., Ndongo, M., Mbodj, B., Coundoul, B., Gueye, P. A. T., Seck, M. C., & Seck, S. M. (2026). Performance of point-of-care creatinine testing in the estimation of glomerular filtration rate: comparison with laboratory creatinine and cystatin C-based measurements. Journal of Nephrology. https://doi.org/10.1093/joneph/aajag080

Vancouver

Sarr IL, Faye C, Ndongo M, Mbodj B, Coundoul B, Gueye PAT, et al. Performance of point-of-care creatinine testing in the estimation of glomerular filtration rate: comparison with laboratory creatinine and cystatin C-based measurements. Journal of Nephrology. 2026. doi:10.1093/joneph/aajag080

BibTeX
@article{Sarr2026,
  author = {Sarr, Ibrahima Lyra and Faye, Cheikh and Ndongo, Modou and Mbodj, Babacar and Coundoul, Baratou and Gueye, Papa Aly Thiam and Seck, Mame Cheikh and Seck, Sidy Mohamed},
  title = {Performance of point-of-care creatinine testing in the estimation of glomerular filtration rate: comparison with laboratory creatinine and cystatin C-based measurements},
  journal = {Journal of Nephrology},
  year = {2026},
  doi = {10.1093/joneph/aajag080}
}
Kostenlos starten →