Multielectrode catheter-based pulsed electric field vs. cryoballoon for atrial fibrillation ablation: a systematic review and meta-analysis
Europace · 03.12.2024 · 26 (12)
DOI: 10.1093/europace/euae293 ↗Kurz gefasst
Systematische Übersicht in der Zeitschrift Europace, online erschienen am 23.11.2024. Verfasst von Giampaolo Vetta und 26 weiteren Autor:innen. Beteiligte Einrichtung: Heart Rhythm Management Centre, Postgraduate Program in Cardiac Electrophysiology and Pacing, Universitair Ziekenhuis Brussel-Vrije Universiteit Brus…. Hauptthemen laut MEDLINE: Atrial Fibrillation, Cryosurgery, Catheter Ablation. Die Arbeit wurde bisher 21-mal zitiert (Crossref) und nennt 28 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Aims
Pulsed field ablation (PFA) is an innovative technology recently adopted for the treatment of atrial fibrillation (AF). Preclinical and clinical studies have reported a remarkable safety profile, as a result of its tissue-specific effect targeting cardiomyocytes and sparing adjacent tissues. Single-shot pentaspline system was the first PFA device to receive regulatory approval. We performed a meta-analysis to compare the efficacy and safety of PFA with the single-shot pentaspline system vs. currently available second-/third-/fourth-generation cryoballoon ablation (CRYO) technologies. Methods and results We systematically searched electronic databases for studies focusing on AF ablation employing the PFA single-shot pentaspline system or second-/third-/fourth-generation CRYO technologies. The primary endpoints were acute procedural success assessed on a vein and patient basis. Safety endpoints included overall periprocedural complications and major periprocedural complications. We also compared procedural, fluoroscopy times, and freedom from atrial tachyarrhythmias (ATs) at follow-up (secondary endpoints). Twenty and 70 studies were included for PFA and CRYO, respectively. Pulsed field ablation demonstrated greater acute procedural success on a vein basis (99.9% vs. 99.1%; P < 0.001), as well as per patient (99.5% vs. 98.4%; P < 0.001). Pulsed field ablation yielded lower overall periprocedural complications (3.1% vs. 5.6%; P < 0.001), shorter procedural time (75.9 min vs. 105.6 min; P < 0.001), and fluoroscopy time (14.2 min vs. 18.9 min; P < 0.001) compared with CRYO. No differences were found for major periprocedural complications (1.2% vs. 1.0%; P = 0.46) and freedom from ATs at 1 year (82.3% vs. 80.3%; log-rank P = 0.61).
Conclusion
Pulsed field ablation contributed to higher acute procedural success and safety compared with CRYO. No statistically significant differences in AT recurrence at 1-year follow-up were observed.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Heart Rhythm Management Centre, Postgraduate Program in Cardiac Electrophysiology and Pacing, Universitair Ziekenhuis Brussel-Vrije Universiteit Brussel, European Reference Networks Guard-Heart, Brussels 1090, Belgium.
- Mediterranean Consortium for Arrhythmia Research (MediCAR), Rome, Italy.
- Texas Cardiac Arrhythmia Institute, St. David's Medical Center, Austin, TX 78705, USA.
- Department of Clinical and Experimental Medicine, Cardiology Unit, University of Messina, Messina, Italy.
- Arrhythmology Unit, Ospedale Fatebenefratelli Isola Tiberina-Gemelli Isola, Rome, Italy.
- Heart Rhythm Department, Clinique Pasteur, Toulouse, France.
ORCID-Profile: Giampaolo Vetta, Domenico Giovanni Della Rocca, Antonio Parlavecchio, Michele Magnocavallo, Antonio Sorgente, Luigi Pannone, Alvise Del Monte, Alexandre Almorad, Lorenzo Marcon, Ioannis Doundoulakis, Sanghamitra Mohanty, Charles Audiat, Kazutaka Nakasone, Gezim Bala, Erwin Ströker, Stéphane Combes, Ingrid Overeinder, Stefano Bianchi, Pietro Palmisano, Pietro Rossi, Serge Boveda, Andrea Natale, Andrea Sarkozy, Carlo de Asmundis, Gian-Battista Chierchia
Zitieren
APA
Vetta, G., Della Rocca, D. G., Parlavecchio, A., Magnocavallo, M., Sorgente, A., Pannone, L., Del Monte, A., Almorad, A., Sieira, J., Marcon, L., Doundoulakis, I., Mohanty, S., Audiat, C., Nakasone, K., Bala, G., Ströker, E., Combes, S., Overeinder, I., Bianchi, S., …, & Chierchia, G. B. (2024). Multielectrode catheter-based pulsed electric field vs. cryoballoon for atrial fibrillation ablation: a systematic review and meta-analysis. Europace, 26(12). https://doi.org/10.1093/europace/euae293
Vancouver
Vetta G, Della Rocca DG, Parlavecchio A, Magnocavallo M, Sorgente A, Pannone L, et al. Multielectrode catheter-based pulsed electric field vs. cryoballoon for atrial fibrillation ablation: a systematic review and meta-analysis. Europace. 2024;26(12). doi:10.1093/europace/euae293
BibTeX
@article{Vetta2024,
author = {Vetta, Giampaolo and Della Rocca, Domenico Giovanni and Parlavecchio, Antonio and Magnocavallo, Michele and Sorgente, Antonio and Pannone, Luigi and Del Monte, Alvise and Almorad, Alexandre and Sieira, Juan and Marcon, Lorenzo and Doundoulakis, Ioannis and Mohanty, Sanghamitra and Audiat, Charles and Nakasone, Kazutaka and Bala, Gezim and Ströker, Erwin and Combes, Stéphane and Overeinder, Ingrid and Bianchi, Stefano and Palmisano, Pietro and Rossi, Pietro and Boveda, Serge and La Meir, Marc and Natale, Andrea and Sarkozy, Andrea and de Asmundis, Carlo and Chierchia, Gian-Battista},
title = {Multielectrode catheter-based pulsed electric field vs. cryoballoon for atrial fibrillation ablation: a systematic review and meta-analysis},
journal = {Europace},
year = {2024},
volume = {26},
number = {12},
doi = {10.1093/europace/euae293}
}
EUVIMED