Valve Expectoration After Endobronchial Valve Placement: Potential Utility of Computed Tomography‐Based Airway Sizing—A Case Report
Respirology Case Reports · 2026-11 · 14 (11)
DOI: 10.1002/rcr2.70817 ↗Abstract
ABSTRACT A 75‐year‐old man with chronic obstructive pulmonary disease underwent endobronchial valve (EBV) placement in the left lower lobe and expectorated the valve from left B10 1 month later. Retrospective analysis of preprocedural dynamic computed tomography (CT) images showed that the median B10 luminal diameter increased from 4.25 mm at end‐expiration to 6.55 mm at end‐inspiration, approaching the upper limit of the recommended working range (4.0–7.0 mm) for the initially placed size 4.0 valve. A size 5.5 valve was subsequently placed without further expectoration. Bronchoscopic sizing under general anaesthesia with neuromuscular blockade and positive‐pressure ventilation may underestimate the maximum airway calibre encountered during spontaneous breathing, particularly in dependent bronchi. Inspiratory CT‐based airway assessment may provide useful complementary information for planning EBV placement and selecting an appropriate valve size.