Enhanced 3D Filter‐Exchange Imaging Using Multi‐Slab Acquisition
Magnetic Resonance in Medicine · 2027-01 · 97 (1) · 269-275
DOI: 10.1002/mrm.70593 ↗Abstract
ABSTRACT Purpose Starting from the implementation of a single‐slice FEXI sequence, a 3D method was developed to accelerate a FEXI experiment and acquire a multi‐slice volume. Methods 3D multi‐slab FEXI divides the volume of interest into different slabs. In each slab, the slab‐selective diffusion and exchange modules are combined with slice‐selective detection modules. The 3D multi‐slab FEXI acquisition speed was assessed against single‐ and multi‐slice FEXI. In addition, the new method was compared to single‐slice FEXI on a yeast phantom. Results 3D multi‐slab FEXI acquires thirty slices within the same scan time required for single‐ and multi‐slice FEXI to acquire one and five slices, respectively. 3D multi‐slab FEXI found AXR values, homogeneity, repeatability, and reproducibility comparable to those of single‐slice FEXI, increasing SNR efficiency. Conclusion 3D multi‐slab FEXI accelerates a FEXI experiment and acquires a volume of several slices in the same scan time as the single‐slice FEXI protocol, giving comparable AXRs.
EUVIMED