Echo‐Planar‐Based Time‐of‐Flight Imaging Using a Modified Interleaved Flyback Trajectory
Magnetic Resonance in Medicine · 2026
DOI: 10.1002/mrm.70272 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2026. Verfasst von Simon Blömer und 2 weiteren Autor:innen. Beteiligte Einrichtung: German Center for Neurodegenerative Diseases (DZNE), Bonn, Germany. Hauptthemen laut MEDLINE: Echo-Planar Imaging, Image Enhancement, Imaging, Three-Dimensional, Image Interpretation, Computer-Assisted. Die Arbeit nennt 40 Literaturangaben. Gefördert von: Helmholtz Association, Initiative and Networking Fund.
Abstract (Originalsprache)
Purpose
A modified interleaved flyback (miFB) approach is introduced, designed to mitigate flow artifacts caused by alternating readout polarities in Echo Planar Imaging (EPI), while preserving acquisition efficiency.
Methods
We propose reconstructing odd and even echoes of 3D‐EPI separately. To this end, the respective missing lines are acquired in interleaved shots with inverted polarity and an additional gradient pre‐lobe. Thereby, high scan efficiency is maintained compared to unsampled flyback gradients. Our miFB approach is additionally combined with gradient moment smoothing and compared to the interleaved dual‐echo with acceleration (IDEA) method in phantom and in vivo scans at 7 Tesla.
Results
Our results demonstrate a significant reduction in ghosting and signal dropout using the miFB approach, yielding comparable image quality to non‐EPI acquisitions while reducing acquisition time by approximately half.
Conclusion
The miFB approach offers a substantial reduction in flow artifacts, allowing for decreased acquisition times in TOF‐MRA.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- German Center for Neurodegenerative Diseases (DZNE), Bonn, Germany.
- Department of Physics and Astronomy, University of Bonn, Bonn, Germany.
ORCID-Profile: Simon Blömer, Tony Stöcker, Rüdiger Stirnberg
Förderung
- Helmholtz Association, Initiative and Networking Fund
Zitieren
APA
Blömer, S., Stöcker, T., & Stirnberg, R. (2026). Echo‐Planar‐Based Time‐of‐Flight Imaging Using a Modified Interleaved Flyback Trajectory. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.70272
Vancouver
Blömer S, Stöcker T, Stirnberg R. Echo‐Planar‐Based Time‐of‐Flight Imaging Using a Modified Interleaved Flyback Trajectory. Magnetic Resonance in Medicine. 2026. doi:10.1002/mrm.70272
BibTeX
@article{Blmer2026,
author = {Blömer, Simon and Stöcker, Tony and Stirnberg, Rüdiger},
title = {Echo‐Planar‐Based Time‐of‐Flight Imaging Using a Modified Interleaved Flyback Trajectory},
journal = {Magnetic Resonance in Medicine},
year = {2026},
doi = {10.1002/mrm.70272}
}
EUVIMED