Saturation Power Dispersion Analysis for the Imaging of Amide Proton Transfer ( APT ) and Nuclear Overhauser Effect ( NOE )

Yifan Zhao iD, Tao Jin iD, Bistra Iordanova iD

Magnetic Resonance in Medicine · 2026

DOI: 10.1002/mrm.70151 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2026. Verfasst von Yifan Zhao und 2 weiteren Autor:innen. Beteiligte Einrichtung: Department of Bioengineering, University of Pittsburgh, Pittsburgh, Pennsylvania, USA. Hauptthemen laut MEDLINE: Magnetic Resonance Imaging, Brain, Amides, Stroke. Die Arbeit nennt 59 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: NINDS NIH HHS, NIH HHS, National Institutes of Health.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

APT and NOE MRI signals are usually obtained from a Z‐spectrum, but the separation of background relaxation effects has certain limitations. We propose to obtain APT and NOE signals from a power dispersion spectrum.

Methods

Theoretical analysis showed that in the power dispersion spectrum of the apparent relaxation ratio (ARR), APT and NOE signals can be separated by a double‐Lorentzian model. Simulation, phantom, and in vivo rodent stroke experiments were performed to evaluate the feasibility of this method.

Results

Simulation and phantom results showed that the ARR of various signal contributions are Lorentzians with different magnitudes and linewidths, and the linewidth of the Lorentzian from APT or NOE is much narrower than those from other relaxation effects. In rat brain, the power dispersion spectra were acquired in the 0.3–4.2 μT range, where the ARR for APT at 3.6 ppm and for NOE at −3.6 ppm can both be effectively approximated by a double Lorentzian model. The Lorentzian linewidth for APT and NOE was 18% and 20% of that from the total background relaxations, respectively. Compared to normal tissue, the ischemic brain tissue showed significantly lower ARR amplitude for APT but not for NOE, and there was minimal change in the ARR linewidth for both APT and NOE.

Conclusion

We demonstrated that APT and NOE signals can be obtained from the analysis of the power dispersion spectrum, which does not require a distinct signal peak in the Z‐spectrum and can be a useful tool at lower magnetic fields.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Magnetic Resonance ImagingBrainAmidesStrokeImage Processing, Computer-AssistedAnimalsRatsPhantoms, ImagingProtonsAlgorithmsComputer SimulationRats, Sprague-DawleyMale

Stichwörter der Autor:innen

APTLorentzianNOEapparent relaxation ratiolinewidthpower dispersion

Substanzen

ProtonsAmides

Autor:innen und Institutionen

  • Department of Bioengineering, University of Pittsburgh, Pittsburgh, Pennsylvania, USA.
  • Department of Radiology, University of Pittsburgh, Pittsburgh, Pennsylvania, USA.

ORCID-Profile: Yifan Zhao, Tao Jin, Bistra Iordanova

Förderung

  • NINDS NIH HHS
  • NIH HHS
  • National Institutes of Health

Zitieren

APA

Zhao, Y., Jin, T., & Iordanova, B. (2026). Saturation Power Dispersion Analysis for the Imaging of Amide Proton Transfer ( APT ) and Nuclear Overhauser Effect ( NOE ). Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.70151

Vancouver

Zhao Y, Jin T, Iordanova B. Saturation Power Dispersion Analysis for the Imaging of Amide Proton Transfer ( APT ) and Nuclear Overhauser Effect ( NOE ). Magnetic Resonance in Medicine. 2026. doi:10.1002/mrm.70151

BibTeX
@article{Zhao2026,
  author = {Zhao, Yifan and Jin, Tao and Iordanova, Bistra},
  title = {Saturation Power Dispersion Analysis for the Imaging of Amide Proton Transfer ( APT ) and Nuclear Overhauser Effect ( NOE )},
  journal = {Magnetic Resonance in Medicine},
  year = {2026},
  doi = {10.1002/mrm.70151}
}
Kostenlos starten →