Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2025. Verfasst von Dongyue Si und 9 weiteren Autor:innen. Beteiligte Einrichtung: School of Biomedical Engineering and Imaging Sciences, King's College London, London, UK. Hauptthemen laut MEDLINE: Heart, Magnetic Resonance Imaging, Image Processing, Computer-Assisted, Image Interpretation, Computer-Assisted. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 33 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: EPSRC, NHLBI NIH HHS, Millenium Institute for Intelligent Healthcare Engineering.
Abstract (Originalsprache)
Purpose
To propose and validate a cardiac T 1 ρ mapping sequence at 0.55 T comparing continuous‐wave and adiabatic spin‐lock (SL) preparation pulses.
Methods
The proposed 2D sequence acquires four single‐shot balanced SSFP readout images with differing contrasts in a single breath‐hold. The first three images are prepared with T 1 ρ preparation pulses with different durations, while the last image uses a saturation pulse immediately before data acquisition. The T 1 ρ map is calculated using a 3‐parameter fitting method. Bloch equation simulations were performed to optimize the parameters of the adiabatic‐SL pulses. Phantom studies and in vivo experiments in 10 healthy volunteers, a porcine myocardial infarction model, and a patient with suspected hypertrophic cardiomyopathy were performed to validate the performance of the proposed adiabatic T 1 ρ (T 1 ρ Ad ) mapping in comparison with conventional continuous‐wave T 1 ρ (T 1 ρ CW ) mapping.
Results
The adiabatic‐SL pulse with simulation‐optimized parameters demonstrated robust performance despite B 0 and B 1 field inhomogeneities. Phantom T 1 ρ CW and T 1 ρ Ad mapping exhibited comparable precision. In vivo experiments on healthy volunteers showed that myocardial T 1 ρ Ad is higher than T 1 ρ CW (106.1 ± 7.1 vs. 47.0 ± 5.1 ms, p < 0.01) with better precision (11.4% ± 2.6% vs. 14.5% ± 2.1%, p < 0.01) and less spatial variation (10.9% ± 3.0% vs. 14.4% ± 3.4%, p < 0.01). Both T 1 ρ CW and T 1 ρ Ad mapping agreed with late gadolinium enhancement findings in the porcine model and the patient, and exhibited improved contrast compared to T 1 and T 2 mapping.
Conclusion
Both T 1 ρ CW and T 1 ρ Ad are promising for non‐contrast detection of various cardiomyopathies at 0.55 T, but T 1 ρ Ad demonstrates better spatial uniformity than T 1 ρ CW .
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- School of Biomedical Engineering and Imaging Sciences, King's College London, London, UK.
- MR Research Collaborations, Siemens Healthcare Limited, Camberley, UK.
- Department of Biomedical Engineering, The Ohio State University, Columbus, Ohio, USA.
- Dorothy M. Davis Heart and Lung Research Institute, The Ohio State University, Columbus, Ohio, USA.
- Department of Emergency Medicine, The Ohio State University, Columbus, Ohio, USA.
- Division of Cardiovascular Medicine, Department of Internal Medicine, The Ohio State University, Columbus, Ohio, USA.
ORCID-Profile: Dongyue Si, Michael G Crabb, Orlando P Simonetti
Förderung
- EPSRC
- NHLBI NIH HHS
- Millenium Institute for Intelligent Healthcare Engineering
- the Department of Health through the National Institute for Health Research (NIHR), Cardiovascular MedTech Co-operative, Comprehensive Biomedical Research Cent…
- BHF
- IMPACT, Center of Interventional Medicine for Precision and Advanced Cellul
- National Heart, Lung, and Blood Institute of the National Institutes of Health
- Wellcome Trust
Zitieren
APA
Si, D., Crabb, M. G., Littlewood, S. J., Kunze, K. P., Varghese, J., Binzel, K., Khan, M., Simonetti, O. P., Prieto, C., & Botnar, R. M. (2025). Initial experience of cardiac T 1 ρ mapping at 0.55 T : Continuous wave versus adiabatic spin‐lock preparation pulses. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.30582
Vancouver
Si D, Crabb MG, Littlewood SJ, Kunze KP, Varghese J, Binzel K, et al. Initial experience of cardiac T 1 ρ mapping at 0.55 T : Continuous wave versus adiabatic spin‐lock preparation pulses. Magnetic Resonance in Medicine. 2025. doi:10.1002/mrm.30582
BibTeX
@article{Si2025,
author = {Si, Dongyue and Crabb, Michael G and Littlewood, Simon J and Kunze, Karl P and Varghese, Juliet and Binzel, Katherine and Khan, Mahmood and Simonetti, Orlando P and Prieto, Claudia and Botnar, René M},
title = {Initial experience of cardiac T 1 ρ mapping at 0.55 T : Continuous wave versus adiabatic spin‐lock preparation pulses},
journal = {Magnetic Resonance in Medicine},
year = {2025},
doi = {10.1002/mrm.30582}
}
EUVIMED