EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Free‐running 5D whole‐heart MRI for isotropic cardiac function measurements at 3T without contrast agents

Augustin C Ogier iD, Isabel Montón Quesada iD, Xavier Sieber iD, Pauline Calarnou iD, Jean-Baptiste Ledoux iD, Bastien Milani iD, Panagiotis Antiochos iD, Juerg Schwitter iD, Christopher W Roy iD, Jérôme Yerly iD, Matthias Stuber iD, Ruud B van Heeswijk iD

Magnetic Resonance in Medicine · 2025

DOI: 10.1002/mrm.30469 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2025. Verfasst von Augustin C Ogier und 11 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology, Lausanne University Hospital (CHUV) and University of Lausanne (UNIL), Lausanne, Switzerland. Hauptthemen laut MEDLINE: Heart, Magnetic Resonance Imaging, Cine, Imaging, Three-Dimensional, Image Processing, Computer-Assisted. Die Arbeit wurde bisher 13-mal zitiert (Crossref) und nennt 46 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Swiss Heart Foundation, Swiss National Science Foundation, Schweizerischer Nationalfonds zur Förderung der Wissenschaftlichen Forschung.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

To optimize and characterize an interrupted 5D free‐running framework at 3 T for detailed cardiac function assessment without the use of breath holding or contrast agents.

Methods

A free‐running 3D radial gradient echo sequence was periodically interrupted with a preparation and a recovery module to optimize native blood‐to‐myocardium contrast at 3 T. Lipid signal was suppressed using a numerically optimized water‐excitation RF pulse to reduce lipid streaking artifacts and to improve overall image quality. Optimal acquisition parameters were established for a 5‐min scan time using extended phase graph simulations. A compressed sensing‐based reconstruction incorporating cardiac and respiratory inter‐bin deformation fields was employed to generate 5D images of the whole heart. The sharpness and contrast between the left ventricular blood pool and myocardium, along with the functional measurements of the left ventricle from the 5D datasets, were compared to routine 2D cine imaging in 16 healthy volunteers and three patients referred for clinically indicated CMR.

Results

The proposed method resulted in lower contrast vs. and sharpness vs. , but enabled similar left‐ventricle ejection fraction assessment , limits of , intraclass correlation with high reproducibility compared to 2D cine.

Conclusion

The proposed contrast‐free, interrupted free‐running 5D imaging provides left ventricular functional assessments comparable to 2D cine at 3 T, while offering an improved patient experience through shorter scan times and free breathing.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

HeartMagnetic Resonance Imaging, CineImaging, Three-DimensionalImage Processing, Computer-AssistedMagnetic Resonance ImagingHumansAdultContrast MediaMaleFemaleMiddle AgedArtifactsAlgorithmsHeart VentriclesHealthy VolunteersImage Interpretation, Computer-Assisted

Stichwörter der Autor:innen

T 2 $$ {\mathrm{T}}_2 $$ preparation pulse3T5D whole‐heart MRIcompressed sensingcontrast‐agent‐freefree‐runningnon‐rigid motion correction

Substanzen

Contrast Media

Autor:innen und Institutionen

  • Department of Radiology, Lausanne University Hospital (CHUV) and University of Lausanne (UNIL), Lausanne, Switzerland.
  • CIBM Center for Biomedical Imaging (CIBM), Lausanne, Switzerland.
  • Heart and Vessel Department, Lausanne University Hospital (CHUV) and University of Lausanne (UNIL), Lausanne, Switzerland.

ORCID-Profile: Augustin C Ogier, Isabel Montón Quesada, Xavier Sieber, Pauline Calarnou, Jean-Baptiste Ledoux, Bastien Milani, Panagiotis Antiochos, Juerg Schwitter, Christopher W Roy, Jérôme Yerly, Matthias Stuber, Ruud B van Heeswijk

Förderung

  • Swiss Heart Foundation
  • Swiss National Science Foundation
  • Schweizerischer Nationalfonds zur Förderung der Wissenschaftlichen Forschung
  • Schweizerische Herzstiftung

Zitieren

APA

Ogier, A. C., Montón Quesada, I., Sieber, X., Calarnou, P., Ledoux, J. B., Milani, B., Antiochos, P., Schwitter, J., Roy, C. W., Yerly, J., Stuber, M., & van Heeswijk, R. B. (2025). Free‐running 5D whole‐heart MRI for isotropic cardiac function measurements at 3T without contrast agents. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.30469

Vancouver

Ogier AC, Montón Quesada I, Sieber X, Calarnou P, Ledoux JB, Milani B, et al. Free‐running 5D whole‐heart MRI for isotropic cardiac function measurements at 3T without contrast agents. Magnetic Resonance in Medicine. 2025. doi:10.1002/mrm.30469

BibTeX
@article{Ogier2025,
  author = {Ogier, Augustin C and Montón Quesada, Isabel and Sieber, Xavier and Calarnou, Pauline and Ledoux, Jean-Baptiste and Milani, Bastien and Antiochos, Panagiotis and Schwitter, Juerg and Roy, Christopher W and Yerly, Jérôme and Stuber, Matthias and van Heeswijk, Ruud B},
  title = {Free‐running 5D whole‐heart MRI for isotropic cardiac function measurements at 3T without contrast agents},
  journal = {Magnetic Resonance in Medicine},
  year = {2025},
  doi = {10.1002/mrm.30469}
}
Kostenlos starten →