In vivo T2 measurements of the fetal brain using single‐shot fast spin echo sequences
Magnetic Resonance in Medicine · 2024
DOI: 10.1002/mrm.30094 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2024. Verfasst von Suryava Bhattacharya und 11 weiteren Autor:innen. Beteiligte Einrichtung: School of Biomedical Engineering and Imaging Sciences, King's College London, London, UK. Hauptthemen laut MEDLINE: Phantoms, Imaging, Brain, Magnetic Resonance Imaging, Fetus. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 66 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Wellcome Trust, National Institute for Health Research Clinical Research Facility, Wellcome/EPSRC Centre for Medical Engineering.
Abstract (Originalsprache)
Purpose
We propose a quantitative framework for motion‐corrected T2 fetal brain measurements in vivo and validate the single‐shot fast spin echo (SS‐FSE) sequence to perform these measurements.
Methods
Stacks of two‐dimensional SS‐FSE slices are acquired with different echo times (TE) and motion‐corrected with slice‐to‐volume reconstruction (SVR). The quantitative T2 maps are obtained by a fit to a dictionary of simulated signals. The sequence is selected using simulated experiments on a numerical phantom and validated on a physical phantom scanned on a 1.5T system. In vivo quantitative T2 maps are obtained for five fetuses with gestational ages (GA) 21–35 weeks on the same 1.5T system.
Results
The simulated experiments suggested that a TE of 400 ms combined with the clinically utilized TEs of 80 and 180 ms were most suitable for T2 measurements in the fetal brain. The validation on the physical phantom confirmed that the SS‐FSE T2 measurements match the gold standard multi‐echo spin echo measurements. We measured average T2s of around 200 and 280 ms in the fetal brain grey and white matter, respectively. This was slightly higher than fetal T2* and the neonatal T2 obtained from previous studies.
Conclusion
The motion‐corrected SS‐FSE acquisitions with varying TEs offer a promising practical framework for quantitative T2 measurements of the moving fetus.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- School of Biomedical Engineering and Imaging Sciences, King's College London, London, UK.
- Guy's and St Thomas' NHS Foundation Trust, London, UK.
- Centre for the Developing Brain, King's College London, London, UK.
- Biomedical Image Technologies, ETSI Telecomunicración, Universidad Politécnica de Madrid and CIBER-BBN, Madrid, Spain.
ORCID-Profile: Suryava Bhattacharya, Anthony N Price, Alena Uus, Lucilio Cordero-Grande, Rui Pedro A G Teixeira, Shaihan J Malik, Maria Deprez
Förderung
- Wellcome Trust
- National Institute for Health Research Clinical Research Facility
- Wellcome/EPSRC Centre for Medical Engineering
- EPSRC Centre for Doctoral Training in Medical Imaging
- Wellcome EPSRC Centre for Medical Engineering
Zitieren
APA
Bhattacharya, S., Price, A. N., Uus, A., Sousa, H. S., Marenzana, M., Colford, K., Murkin, P., Lee, M., Cordero-Grande, L., Teixeira, R. P. A. G., Malik, S. J., & Deprez, M. (2024). In vivo T2 measurements of the fetal brain using single‐shot fast spin echo sequences. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.30094
Vancouver
Bhattacharya S, Price AN, Uus A, Sousa HS, Marenzana M, Colford K, et al. In vivo T2 measurements of the fetal brain using single‐shot fast spin echo sequences. Magnetic Resonance in Medicine. 2024. doi:10.1002/mrm.30094
BibTeX
@article{Bhattacharya2024,
author = {Bhattacharya, Suryava and Price, Anthony N and Uus, Alena and Sousa, Helena S and Marenzana, Massimo and Colford, Kathleen and Murkin, Peter and Lee, Maggie and Cordero-Grande, Lucilio and Teixeira, Rui Pedro A G and Malik, Shaihan J and Deprez, Maria},
title = {In vivo T2 measurements of the fetal brain using single‐shot fast spin echo sequences},
journal = {Magnetic Resonance in Medicine},
year = {2024},
doi = {10.1002/mrm.30094}
}
EUVIMED