EUVIMEDEUVIMED entdecken
Open Access · CC BY-NC-ND 4.0

Universal modes: Calibration‐free time‐interleaved acquisition of modes

Simon Schmidt iD, Xiaoxuan He iD, Gregory J Metzger iD

Magnetic Resonance in Medicine · 2024

DOI: 10.1002/mrm.30032 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2024. Verfasst von Simon Schmidt und 2 weiteren Autor:innen. Beteiligte Einrichtung: Center for Magnetic Resonance Research, University of Minnesota, Minneapolis, Minnesota, USA. Hauptthemen laut MEDLINE: Heart, Image Processing, Computer-Assisted, Algorithms, Pelvis. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 31 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: NIBIB NIH HHS, NCI NIH HHS, National Institute of Biomedical Imaging and Bioengineering.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

To introduce universal modes by applying the universal pulse concept to time‐interleaved acquisition of modes (TIAMO), thereby achieving calibration‐free inhomogeneity mitigation for body imaging at ultra‐high fields.

Methods

Two databases of different RF arrays were used to demonstrate the feasibility of universal modes. The first comprised 31 cardiac in vivo data sets acquired at 7T while the second consisted of 6 simulated 10.5T pelvic data sets. Subject‐specific solutions and universal modes were computed and subsequently evaluated alongside predefined default modes. For the cardiac database, subdivision into subpopulations was investigated. The optimization was performed using least‐squares (LS) TIAMO and acquisition modes optimized for refocused echoes (AMORE). Finally, universal modes based on simulated pelvis data were applied in vivo at 10.5T.

Results

In all studied cases, the universal modes yield improvements over the predefined default modes of up to 51% (cardiac) and 30% (pelvic) in terms of median excitation error when using two modes. The subpopulation‐specific cardiac solutions revealed a further improvement of universal modes at the expense of increased errors when applied outside the appropriate subpopulation. Direct application of simulation‐based universal modes in vivo resulted in up to a 14% reduction in excitation error compared to default modes and up to a 34% reduction in peak 10 g local specific absorption rate (SAR) compared to subject‐specific solutions.

Conclusions

Universal modes are feasible for calibration‐free inhomogeneity mitigation at ultra‐high fields. In addition, simulation‐based solutions can be applied directly in vivo, eliminating the need for large in vivo databases.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

HeartImage Processing, Computer-AssistedAlgorithmsPelvisMagnetic Resonance ImagingHumansCalibrationMaleAdultComputer SimulationPhantoms, ImagingFemaleLeast-Squares AnalysisDatabases, Factual

Stichwörter der Autor:innen

AMOREMRIRF shimmingTIAMObodyparallel transmissionultra‐high field

Autor:innen und Institutionen

  • Center for Magnetic Resonance Research, University of Minnesota, Minneapolis, Minnesota, USA.

ORCID-Profile: Simon Schmidt, Xiaoxuan He, Gregory J Metzger

Förderung

  • NIBIB NIH HHS
  • NCI NIH HHS
  • National Institute of Biomedical Imaging and Bioengineering

Zitieren

APA

Schmidt, S., He, X., & Metzger, G. J. (2024). Universal modes: Calibration‐free time‐interleaved acquisition of modes. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.30032

Vancouver

Schmidt S, He X, Metzger GJ. Universal modes: Calibration‐free time‐interleaved acquisition of modes. Magnetic Resonance in Medicine. 2024. doi:10.1002/mrm.30032

BibTeX
@article{Schmidt2024,
  author = {Schmidt, Simon and He, Xiaoxuan and Metzger, Gregory J},
  title = {Universal modes: Calibration‐free time‐interleaved acquisition of modes},
  journal = {Magnetic Resonance in Medicine},
  year = {2024},
  doi = {10.1002/mrm.30032}
}
Kostenlos starten →