Fetal MRI : Radiofrequency Safety Assessment at 3 Tesla

Filiz Yetisir iD, Esra Abaci Turk, Henry A Feldman, Borjan Gagoski iD, Ryne A Didier, Carol Barnewolt, Judy A Estroff, Lawrence L Wald iD, Elfar Adalsteinsson, P Ellen Grant

Journal of Magnetic Resonance Imaging · 2025-09 · 62 (3) · 844-853

DOI: 10.1002/jmri.29797 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Magnetic Resonance Imaging, online erschienen am 17.04.2025. Verfasst von Filiz Yetisir und 9 weiteren Autor:innen. Beteiligte Einrichtung: Department of Pediatrics, Harvard Medical School, Boston, Massachusetts, USA. Hauptthemen laut MEDLINE: Fetus, Magnetic Resonance Imaging. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 36 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: NICHD NIH HHS, CSR NIH HHS, NIBIB NIH HHS.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

3‐T MRI can improve image quality of fetal imaging compared to 1.5‐T MRI. However, concerns exist regarding increased local tissue heating at 3‐T.

Purpose

To assess fetal MRI radiofrequency (RF) safety at 3‐T by comparing simulated tissue heating to 1.5‐T (using constant RF exposure) and by simulating tissue heating at 3‐T using RF exposures from clinical fetal examinations. Study Type Retrospective. Population Seven voxelized anatomical pregnant body models (gestational age [GA] 30 ± 3 weeks [mean ± standard deviation], maternal body mass index [BMI] 27.8 ± 8.5 kg/m 2 ) were used. Maternal whole‐body average specific absorption rate (wbSAR) logs were collected from 85 clinical examinations at 3‐T (GA 25 ± 6 weeks, BMI 30.3 ± 6.8 kg/m 2 ). Field Strength/Sequence 3‐T, 1.5‐T, HASTE, VIBE, TRUFISP, EPI, DTI. Assessment Simulated maternal and fetal peak and average SAR, temperature, and peak thermal dose were compared at 3‐T and 1.5‐T for 60 min 2 W/kg wbSAR using 7 body models and a 16‐rung band‐pass RF coil. Temperature and thermal dose were simulated in one body model using clinical wbSAR exposures at 3‐T. Statistical Tests Factorial analysis of variance was performed using 28 maternal and fetal temperature measurements from 7 body models to detect a difference between 3‐T and 1.5‐T. p < 0.05 was considered statistically significant.

Results

For constant RF exposure, we found no difference between 3‐T and 1.5‐T in peak maternal (1.5‐T:40.38 ± 0.21°C; 3‐T:40.40 ± 0.20°C; p = 0.85), peak fetal (1.5‐T:39.21 ± 0.17°C; 3‐T:39.09 ± 0.16°C; p = 0.19), and average maternal (1.5‐T:37.32 ± 0.05°C; 3‐T:37.33 ± 0.04°C; p = 0.68) temperature. We observed significantly higher average fetal temperatures at 1.5‐T (1.5‐T:37.75 ± 0.06°C; 3‐T:37.70 ± 0.05°C). For 3‐T clinical RF exposures, simulated peak temperatures exceeded the recommended limits. However, the thermal dose was below the recommended limit. Data Conclusion For the same RF coil geometry, local heating was similar at 3‐T and 1.5‐T for constant RF exposure. Although realistic 3‐T RF exposures could cause peak temperatures above the recommended limits, thermal dose was below the recommended limit. Evidence Level 1. Technical Efficacy Stage 1.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

FetusMagnetic Resonance ImagingHumansPrenatal DiagnosisFemalePregnancyHot TemperatureModels, AnatomicRadio WavesYoung AdultAdultMiddle Aged

Stichwörter der Autor:innen

1.5 tesla3 teslaRF heatingSARfetal MRIsafety

Autor:innen und Institutionen

  • Department of Pediatrics, Harvard Medical School, Boston, Massachusetts, USA.
  • Fetal Neonatal Neuroimaging and Developmental Science Center, Boston Children's Hospital, Boston, Massachusetts, USA.
  • Biostatistics and Research Design Center, Boston Children's Hospital, Boston, Massachusetts, USA.
  • Department of Radiology, Boston Children's Hospital, Harvard Medical School, Boston, Massachusetts, USA.
  • Fetal Care and Surgery Center, Boston Children's Hospital, Boston, Massachusetts, USA.
  • Harvard-MIT Division of Health Sciences and Technology, Massachusetts Institute of Technology, Cambridge, Massachusetts, USA.

ORCID-Profile: Filiz Yetisir, Borjan Gagoski, Lawrence L Wald

Förderung

  • NICHD NIH HHS
  • CSR NIH HHS
  • NIBIB NIH HHS
  • Center for Scientific Review

Zitieren

APA

Yetisir, F., Abaci Turk, E., Feldman, H. A., Gagoski, B., Didier, R. A., Barnewolt, C., Estroff, J. A., Wald, L. L., Adalsteinsson, E., & Grant, P. E. (2025). Fetal MRI : Radiofrequency Safety Assessment at 3 Tesla. Journal of Magnetic Resonance Imaging, 62(3), 844-853. https://doi.org/10.1002/jmri.29797

Vancouver

Yetisir F, Abaci Turk E, Feldman HA, Gagoski B, Didier RA, Barnewolt C, et al. Fetal MRI : Radiofrequency Safety Assessment at 3 Tesla. Journal of Magnetic Resonance Imaging. 2025;62(3):844-853. doi:10.1002/jmri.29797

BibTeX
@article{Yetisir2025,
  author = {Yetisir, Filiz and Abaci Turk, Esra and Feldman, Henry A and Gagoski, Borjan and Didier, Ryne A and Barnewolt, Carol and Estroff, Judy A and Wald, Lawrence L and Adalsteinsson, Elfar and Grant, P Ellen},
  title = {Fetal MRI : Radiofrequency Safety Assessment at 3 Tesla},
  journal = {Journal of Magnetic Resonance Imaging},
  year = {2025},
  volume = {62},
  number = {3},
  pages = {844--853},
  doi = {10.1002/jmri.29797}
}
Kostenlos starten →