Torsional properties of healed canine diaphyseal defects grafted with a fibrillar collagen and hydroxyapatite/tricalcium phosphate composite
Journal of Applied Biomaterials · 1994-12 · 5 (4) · 277-283
DOI: 10.1002/jab.770050402 ↗Abstract
Abstract The need for alternatives to autogenous bone grafts is widely recognized. This study compared the torsional strength of canine femora 1 year after grafting with one of three forms of a collagen/hydroxyapatite/tricalcium phosphate bone grafting material (COLLAGRAFT 15 ), autogenous bone, or no graft. The groups were compared to each other and to the unoperated contralateral femora. Results of torsional testing were evaluated for torsional strength, torsional displacement, total energy to fracture and White fracture mode. Data analysis showed lower torsional strength of the operated vs. unoperated femora with the exception of morsellized COLLAGRAFT 15 material, which had higher strength. However, the only difference in the operated groups was that the morsellized COLLAGRAFT 15 had greater strength than several groups including the autogenous bone group. There was no difference found in angular displacement between any of the groups. However, there was a difference in the energy to fracture in both strip forms of the COLLAGRAFT 15 . The final conclusion is that in this model, grafting with COLLAGRAFT 15 provided torsional properties at one year postoperatively at least equivalent to autogenous bone. © 1994 John Wiley & Sons, Inc.